Digital Two-wire HART Vortex Flowmeter Transmitter Based on Double Charge Amplifier |
SHU Zhang-ping,XU Ke-jun,SHAO Chun-li,FANG Min |
School of Electrical and Automation Engineering, Hefei University of Technology, Hefei, Anhui 230009, China |
|
|
Abstract A digital two wire HART vortex flowmeter transmitter based on double charge amplifiers is developed in order to solve the problems that there is low sensitivity in the high temperature resistant vortex flowmeter sensor, and HART communication increases power consumption and its stability is easily affected by the power. The signal conditioning circuit concluding two-stage charge amplifier and two-stage low-pass filter is designed to amplify and filter the low output amplitude of high temperature resistant vortex flow sensor, and obtain the stable flow signal. Two isolated power supply scheme is adopted to improve load ability of power supply and ensure the stability of the communication, which achieves two-wire HART communication during the digital signal processing. The experimental results show that the transmitter possesses high measurement accuracy, small measurable low limit, and also meets the requirements of power consumption.
|
Received: 22 February 2016
Published: 27 September 2017
|
|
Corresponding Authors:
Ke-Jun Xu
E-mail: dsplab@hfut.edu.cn
|
|
|
|
[1]方敏. 三种数字流量计硬件研制与改进[D]. 合肥:合肥工业大学, 2011.
[2]徐科军, 方敏, 罗清林, 等. 抗振型低功耗数字涡街流量计研制[J]. 计量学报, 2013, 34(5):435-440.
[3]黄云志, 徐科军. 基于小波滤波器组的涡街流量计信号处理方法[J]. 计量学报, 2006, 27(2):133-136.
[4]黄云志, 徐科军. 涡街流量计信号处理方法与系统的研究现状[J]. 自动化仪表, 2003, 24(8):1-4.
[5]Shao C L, Xu K J, Fang M. Frequency-variance based anti-strong vibration interference method for vortex flow sensor[J]. IEEE Trans on Instrumentation and Measurement, 2014, 63(6):1566-1582.
[6]Shao C L, Xu K J, Fang M. Feature patterns extraction-based amplitude/frequency modulation model for for vortex sensor output signal[J]. IEEE on Instrumentation and Measurement, 2015, 64(11):3031-3044.
[7]徐科军, 王沁, 方敏, 等. 基于数字滤波与自相关的涡街流量计抗强振动方法改进[J]. 计量学报, 2014, 35(5):449-454.
[8]徐科军, 王沁, 方敏, 等. 基于噪声模板与互相关计算的涡街流量计抗强振动方法[J]. 计量学报, 2015, 36(1):37-42.
[9]Xu K J, Huang Y Z, LYU X H. Power-spectrum-analysis-based signal processing system for vortexflowmeters[J]. IEEE Trans on IM, 2006, 55(3):1006-1011.
[10]Hondoh M, Wada M, Andoh T, et al. A vortexflowmeter with spectral analysis signal processing[C]//Sensors for Industry, 2001. Proceedings of the First ISA/IEEE Conference , 2001:35-40.10.1109/SFICON.2001.968495.
[11]徐科军, 汪安民, 吕迅竑, 等. 基于自适应陷波的涡街流量计信号处理系统[J]. 计量学报, 2000, 21(3):199-204.
[12]王肖芬, 徐科军, 陈智渊. 基于DSP 的低成本涡街流量计信号处理系统[J]. 仪器仪表学报, 2006, 27(11):1427-1432.
[13]徐科军, 吕迅竑, 陈荣保, 等. 基于DSP 具有谱分析功能的涡街流量计信号处理系统[J]. 仪器仪表学报, 2001, 22( 3):255- 260.
[14]徐科军, 刘三山, 刘家祥, 等.基于数字频谱与模拟带通滤波组的两线制涡街流量计[J].计量学报, 2011, 32(1):25- 30.
[15]方敏, 王沁, 徐科军. 一种新型的电池供电的数字流体振荡式流量计[J]. 计量学报, 2011, 32(6):511- 516.
[16]罗清林, 徐科军, 刘三山. 基于快速傅里叶变换的低功耗两线制涡街流量计[J]. 电子测量与仪器学报, 2010, 24(7):692- 697.
[17]李建平. 基于HART总线的智能型涡街流量计的研究开发[D]. 重庆:重庆大学,2011.
[18]YOKOGAWA. digitalYEWFLO Series Vortex Flowmeter[Z]. 2015. |
|
|
|