Discussion on Improving the Temperature Calibration Capability of Environmental Testing Equipment |
GUO Shen-hui,WU Jia,CAI Ming-xuan,ZHANG Lin |
Qingdao Institute of Measurement Technology, Qingdao, Shandong 266101, China |
|
|
Abstract Discussion on improving environmental test equipment temperature calibration capabilities in several ways, both a higher level of standard device for temperature data logging devices is used to transmit temperature and platinum resistive temperature correction value is used. After using the amount of transfer mode, temperature and humidity meter through the calibration test case, and calculates the temperature calibration uncertainty of results found assess the results of its much better than before, expanded uncertainty is 0.26 ℃ (k = 2). The advantage of this method is full use of existing laboratory measurement equipment, and provides a new idea to improve the environmental test equipment calibration capability .
|
|
|
|
|
|
[1]JJF 1101—2003环境试验设备温度、湿度校准规范[S].
[2]郭沈辉,王林虎,武佳,等.环境温湿度实验设备的校准与不确定度分析[J].工业计量,2012,(S2):76-78. |
|
|
|