Development of a Calibration Device about Tungsten-rhenium Thermocouples in High Temperature |
LÜ Guo-yi,ZHANG He,YANG Yong-jun,LIU Dan-ying,ZHAO Nan |
AVIC Changcheng Institute of Metrology and Measurement, Beijing 100095, China |
|
|
Abstract The composition and the principle of the calibration device about high temperature Tungsten-rhenium thermocouples is introduced. The vertical temperature calibration furnace as heat source, photoelectric pyrometer as standard, to calibrat the Tungsten-rhenium thermocouples. Through using graphite as furnace heating element and carrying on the design of its key structure, it can solve the uniformity of furnace temperature field, fixed connection of the heating element, and the inflation or shrinkage problem in the process of rising or cooling. In order to ensure the radiation temperature measuring accurately, it designs specially the structure of blackbody radiation. Througth analying and designing about the thermocouple protection tube material selection and structural, it solves the pollution problem about thermocouple at high temperature from the heating element volatiling, and adapting to the environment atmosphere in the process of the calibration of thermocouple. The results of experimental and uncertainty evaluation show that the relative uncertainty of temperature calibration results is within 0.6%t in 2 300 ℃, the calibration device can solve technical problems of high temperature thermocouple static traceability.
|
|
|
|
|
|
[1]喬治 E.格拉威,黄美超 译. 耐高温金属热电偶[J].功能材料, 1975,(2):50-55.
[2]王魁汉,吴玉锋,韩元山. 高温热电偶及特殊场合下的温度测量技术[J].世界仪表与自动化, 2003, (3): 42-46.
[3]王魁汉,崔传孟. 抗氧化钨铼热电偶的发展与应用[J].自动化仪表, 1995, 16(6): 1-3.
[4]廖理. 热学计量[M]. 北京: 原子能出版社, 2002.
[5]唐锐. 钨铼热电偶高温分度及其不确定度[J].功能材料, 2004,35(z1):1707-1710.
[6]武建红,赵俭,王松涛. 300~1500℃钨铼热电偶校准结果不确定度分析[J].计测技术,2010,30(s1):71-73. |
|
|
|