Study on Acoustic Volumeter System of Weights |
WANG Xiao-lei, WANG Jian, ZHONG Rui-lin, ZHANG Yue, CAI Chang-qing,YAO Hong, DING Jing-an |
National Institute of Metrology, Beijing 100013, China |
|
|
Abstract To meet the measurement demand of the volume of weights for high accuracy measuring weights, an acoustic volumeter system of weights is developed by using acoustic theory. A practical procedure of measuring the volume of weights ranging from 10 g to 100 g using an acoustic volumeter is proposed. The volumes obtained using an acoustic volumeter were compared with the results obtained by the hydrostatic, and the relative error lower than 8.3×10-4 s acquired. The relative expanded uncertainty of system is lower than 4.3×10-4.
|
|
|
|
|
[1] 姚弘,丁京鞍,王健,等. 砝码的体积测量[J]. 计量技术, 2009, (10): 35-37.
[2] OIML.R 111-1:2004,Weights of classes E1, E2, F1, F2, M1, M1-2, M2, M2-3, and M3 Metrological and technical requirements[S].2004.
[3] TorigocI, Ishii Y. Acoustic bridge volumeter[J]. Trans Soc Instrum Control Eng,2001,E-1(1):164-170.
[4] KobataT, Ueki M, Ooiwa A, et al. Mesurement of the volume of weights using an acoustic volumeter and the reliability of such measurement[J]. Metrologia,2004,41(2):S75-S83.
[5] IshiiY. An acoustic volumeter[J]. Meas Control, 1997, 36: 288-291.
[6] MeasurementScience Laboratory. Instruction Manual of Acoustic Volumeter for Measuring Volume of Weight VM-300[R].2000. |
|
|
|