The Establishment of Radiation Primary Standard in the Range of Low Energy X-rays |
WU Jin-jie1,WANG Pei-wei1,DUAN Xiao-juan1,2,LI Bing1,2,YANG Yuan-di1 |
1.National Institute of Metrology, Beijing 10013, China
2.Chengdu University of Technology, Chengdu, Sichuan 610059, China |
|
|
Abstract The radiation equipment for air-kerma primary standard in the range of low-energy X-rays is established.The results of experiment show that the diameter of the radiation field is more than 40 mm within 1% non-uniformity.The distance from the source to the reference plane is 1 000 mm.Radiation qualities recommend by CCRI are established.The half value layer(HVL)and the homogeneity coefficient are obtained by measuring the curve of x-rays attenuation.This low-energy X-rays radiation equipment and research results are the important base of establishing primary standard,realization of air-kerma and the APMP key comparison in the range of low-energy X-rays(APMP.RI(I)-K2).
|
|
|
|
|
|