|
|
Preliminary Experimental Research on Electron Equilibrium for Calibration of Thermoluminescent Detector |
LI De-hong1,LI Xing-dong1,WU Meng-meng1,2,YANG Yuan-di1 |
1. National Institute of Metrology, Beijing 100013, China
2. Chengdu University of Technology, Chengdu, Sichuan 610051, China |
|
|
Abstract TLD are widely used in environmental radiation monitoring and personal dose monitoring.To ensure the accuracy of TLD calibration in the radiation field, especially when the high-energy γ-rays are used,the front surface of TLD is supposed to be covered by electronic balance material.Through a series of experiments on the PMMA of different thickness as electronic balance material used to be covered on the front surface of TLD in different radiation fields, it is discovered the PMMA of certain thickness must be needed in Co-60 radiation field,while in the Cs-137 radiation field, the PMMA is relative thin.
|
|
|
|
|
|
[1]王寿山.热释光剂量学的研究、现状及发展[J].核电子学与探测技术,1999,19(2): 130-139.
[2]IAEA.National and international standardization of radiation dosimetry[C]// proceedings of an International Symposium on National and International Standardization of Radiation Dosimetry. Atlanta, 1977.
[3]Oberhofer M, Scharmann A.应用热释光剂量学[M].北京:中国计量出版社,1988.
[4]Attix F H,Roesch W C 著,陈常茂译.辐射剂量学[M].北京:原子能出版社,1981.
[5]ISO 4037-1:1996, X and gamma reference radiation for calibrating dosemeters and doserate metersand for determining their response as a function of photon energy-Part 1: Radiation characteristics and production methods[S].
[6]Moralles M,Guimaraes C C,Kuno E O.Response of thermoluminescent dosimeters to photons simulated with the Monte Carlo method [J].Nuclear Instruments and Methods in Physics Research A,2005,545(1-2):261-268. |
|
|
|