The Device of the Neutron Reference Radiation with Radionuclide Neutron Sources |
ZHANG Hui,FAN Cheng,GONG Xiao-ming |
National Institute of Metrology, Beijing 100013, China |
|
|
Abstract The reference radiations produced by radionuclide neutron sources can be used in routine calibration or scaling of the neutron survey meter, it also can be used as the basis platform of some experiments.The neutron reference radiation facility in the National Institute of Metrology is improved according to the ISO 8529—1 and the purposes of the facility.The facility fulfilled with the requirements of routine calibration for neutron instruments, and it can be used for the experiments for evaluating the neutron detectors and neutron protective materials.
|
|
|
|
|
|
[1]汲长松.中子探测实验方法[M].北京:原子能出版社, 1998.
[2]丁大钊,叶春堂,赵志祥,等.中子物理学——原理、方法与应用[M].北京:原子能出版社, 2005.
[3]ISO 8529—1:2001.Reference neutron radiations—Part 1: Characteristics and methods of production[S].
[4]Kluge H.Irradiation facility with radioactive reference neutron sources: Basic principles[R].PTB-N-34,1998.
[5]ISO 8529—2:2000.Calibration fundamentals of radiation protection devices related to the basic quantities characterizing the radiation field[S].
[6]ISO 8529—3:1998. Reference neutron radiations—Part 3:Calibration of area and personal dosimeters and determination of their response as a function of neutron energy and angle of incidence[S].
[7]ISO 4037—1:1996.X and gamma reference radiation for calibrating dosemeters and dose rate meters and determining their response as a function of photon energy—Part 1: Radiation characteristics and production methods[S]. |
|
|
|