Research on the Construction and Reproducibility of the Co-C HTFP for Thermocouple Calibration |
ZHENG Wei1, LIANG Xing-zhong2, WU Jian3 |
1. National Institute of Metrology, Beijing 100013, China;
2. Shandong Institute of Metrology, Jinan, Shandong 250014, China;
3. Beijing Institute of Metrology, Beijing 100029, China |
|
|
Abstract The method of construction of Co-C high temperature fixed point for thermometer calibration and the testing results of reproducibility with thermocouple at National Institute of Metrology of China are introduced.The experimental data show that the reproducibility of melting point of Co-C cell is 0.03 ℃ using Pt/Pt thermocouple, and the cell is robust after 110 melting and freezing cycles with 500 h above 1300 ℃.
|
|
|
|
|
[1] Bloembergen P, Yamada Y, Yamamoto N, et al. Realizing the high-temperature part of a future ITS with the aid of eutectic metal-carbon fixed point[C]//AIP Conference Proceedings,Chicago, Illinois,USA,2003.
[2] Yamad Y, Sasajima N, Sakuma F, et al. Realizing fixed point the copper point up to 2 500 ℃ using metal-carbon eutectics[J].AIST Bulletin of Metrology,2002,1(3): 533-540. [3] Yamada Y. Advances in high-temperature standards above 1 000 ℃[J]. MAPAN-Journal of Metrology Society of India, 2005,20 (2):183-191.
[4] Burns G W, Ripple D C.Techniques for fabrication and annealing Pt/Pd thermocouples for accurate measurements in the Range 0 ℃ to 1 300 ℃[C] // TEMPEMEKO’96,Torino,1996.
[5] Renote E M,Balluello R L, Rusby F E. EUROMET Guidelines for the Construction and Use of Platinum Versus Palladium Thermocouples CCT99-08[S].
[6] Masahiko G.Stability testing of and a reference function for the Pt/Pd thermocouple[C] //TEMPMEKO ’99 , Delft, The Netherlands ,1999.
[7] Edler F. Precise temperature measurement above 1 000 ℃ using thermocouple[J]. Proc Estonian Acad Sci Eng,2007,13(4):310-319.
[8] Yamada Y, Sakuma F, Ono A. Thermocouple observation of melting and Freezing plateaus for metal-carbon eutectic between the copper and palladium points [J]. Metrologia, 2000,37(1):71-73.
[9] Edler F, Zheng W.Investigation of differently designed Pd-C eutectic fixed-point cells for the calibration of thermocouples[J].Metrologia, 2011, 48(5): 275-282.
[10] Burns G W, Ripple D C, Battuello M.Platinum versus palladium thermocouple: an emf-temperature reference function for the Range 0 ℃ to 1 500 ℃[J].Metrologia,1998,35(5):761-780. |
|
|
|