Effect on Optometry for Decentration and Tilt of Lenses in Phoropters |
ZHANG Ji-yan1,2,LIU Wen-li1,SUN Jie1 |
1. Division of Metrology in Ionizing Radiation and Medicine, National Institute of Metrology, Beijing 100013, China
2. Information and Technology School, Beijing Institute of Technology, Beijing 100081, China |
|
|
Abstract With the basic principle of geometric optics and ophthalmic dioptrics, the effect on decentration and tilt of lenses in phoropters is analyzed theoretically and verified by ZEMAX software. Results show that decentration in lenses will introduce extra prismatic power, and tilt will introduce error not only in prismatic power but also in spherical and cylindrical power. Moreover, if front and back surface of one single lens are tilted from two opposite directions, the error will be much bigger than that tilted from the same direction. Besides, the effect of decentration is a little more significant than that of tilt in assembly. Therefore, decentration and tilt from two opposite directions for front and back surface of a lens should be avoided possibly in production and assembly of phoropters.
|
|
|
|
|
|
[1]Bullimore M A, Fusaro R E, Adams C W. The repeatability of automated and clinician refraction[J]. Optometry and Vision Science, 1998, 75(8):617-622.
[2]Strang N C, Winn B, Gray L S, et al. The reliability of subjective refraction[J]. Ophthalmic and Physiological Optics, 1996, 16(3):251.
[3]Li Qi, Ren Qiushi. Correlation among auto-refractor, wavefront aberration, and subjective manual refraction[J]. Proceedings of SPIE, 2005, 5630:159-167.
[4]黄建艳, 陈剑, 周清, 等. 电脑验光仪和综合验光仪联合应用的临床评价[J]. 眼科新进展, 2002, 22(6): 405-407.
[5]李仁芳, 熊斌. 电脑验光、综合验光仪验光及试戴三者度数的比较研究[J]. 赣南医学院学报, 2004, 24(4): 410-412.
[6]Zhang Jiyan, Sun Jie, Gao Mingliang, et al. Metrological research on test method for phoropters[A]. Proceedings of 2007 8th international conference on electronic measurement and instruments[C], Beijing: IEEE Express, 2007, 16-18.
[7]刘文丽,朱建平,洪宝玉,等. 验光仪用柱镜标准器的研制[J]. 计量学报, 2008,30(5): 405-408.
[8]ISO 10341-1997, Ophthalmic instruments-Refractor heads[S].
[9]吴燮灿. 实用眼镜光学[M]. 北京:北京科学技术出版社,2007. |
|
|
|