Research On The Effects Of Bushing On The Triple Point Of Water |
WANG Peng1,2,YAN Xiao-ke2,PU Ling1,ZHANG Jing-yu1,LI Meng1,ZHANG Jia-ren1 |
1. Shanghai Institute of Quality Inspection and Technical Research, Shanghai 201114, China
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract In order to study the effect of bushing on the triple point of water (TPW), the thermometer well which have diameters of 16 mm and 18mm respectively are developed, and the bushing with diameter of 15.5 mm. The resistance of the standard platinum resistance thermometer in the well with bushing and without bushing are measured, by the AC bridge which the accuracy is 0.02×10-6. The experiment results indicated that the bushing have little effect on the value of TPW, less than 30 μK. Additionally, under the same conditions, the more uniformity between the diameters of bushing and thermometer well, the less influence on the values of TPW.
|
Received: 28 October 2014
Published: 10 December 2015
|
|
|
|
|
[1]Thomas H P. The International Temperature Scale of 1990 (ITS-90)[J]. Metrologia, 1990, 27(2):3-10.
[2]杨金环.水三相点自动复现装置的研制及水三相点温度测量的研究[D].北京:北京化工大学,2010.
[3]闫小克,武荷莲,张哲,等.冰桥对水三相点的影响[J].现代测量与实验室管理,2004,(1):25-25.
[4]闫小克.环境对水三相点温度的影响[J].计量技术,2005, (10):31-33.
[5]Nicholas J V, Dransfield T D, Mangum B M. Effects of Heavy Hydrogen and Oxygen on the Triple-Point Temperature of Water[C]// Proceedings of TEMPMEKO, Delft, Netherlands,1999.
[6]Hill K D. Triple point of water cells and the solubility of borosilicate glass [C]// Proceedings of TEMPMEKO,Delft, Netherlands,1999.
[7]Steur P P M, Dematteis R. The use of bushings with triple point of water cells: towards breaking the 50 μK barrier[J]. Metrologia, 2008, 45(5): 529-533.
[8]闫小克,邱萍,王玉兰,等. NIM 水三相点瓶的制作[J]. 计量技术,2003,(1): 61-63.
[9]王鹏,闫小克,葛良全,等.不同计阱的水三相点容器浸没特性的研究[J]. 计量技术,2014, (4): 15-17.
[10]闫小克.液氮冻制冰套法对水三相点温度的影响[J].计量学报,2005,25(4):318-321. |
|
|
|