Research on the Calibration Method of Reference Solar Cell
ZHANG Jun-chao,XIONG Li-min,MENG Hai-feng,HE Ying-wei,ZHANG Bi-feng
National Institute of Metrology, Beijing 100029, China
Download: PDF (322 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  Based on differential spectral responsivity method, a calibration equipment for reference solar cell is established. The absolute responsivity of reference cell varied with temperature of cell and irradiance intensity of bias light are analyzed. The relationship between the short circuit current and the bias current of the solar cell is obtained by numerical simulation techniques. The uncertainty of the result is 1.2% (k=2). Based on outdoor method, a reference cell calibration equipment is established. A detailed uncertainty evaluation of the result is carried out, the value is 2.38% (k=2). Through numerical calculation and analysis of the En value (0.95), it is found that the results measured by the two methods are equivalent.
Key wordsmetrology      reference cell      outdoor calibration      calibration value      responsivity      uncertainty     
Received: 13 May 2015      Published: 28 February 2017
PACS:  TB96  
Corresponding Authors: Jun-chao ZHANG     E-mail: zhangjc@nim.ac.cn
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
ZHANG Jun-chao
XIONG Li-min
MENG Hai-feng
HE Ying-wei
ZHANG Bi-feng
Cite this article:   
ZHANG Jun-chao,XIONG Li-min,MENG Hai-feng, et al. Research on the Calibration Method of Reference Solar Cell[J]. Acta Metrologica Sinica, 2017, 38(2): 171-174.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2017.02.10     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2017/V38/I2/171
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech