|
|
Research on Tube Potential of Industrial X-rays Generator by Invasive Measurement |
WU Jin-jie1,GUO Bin1,LI Bing1,2,JIANG Wei1,2,CHEN Fa-jun1,2,DUAN Sheng-lin1,3 |
1. National Institute of Metrology, Beijing 100029, China;
2. Chengdu University of Technology, Chengdu, Sichuan 610059, China;
3. Beihang University, Beijing 100191, China |
|
|
Abstract An invasive measurement method is established to get the tube potential of industrial X-rays generator base on high voltage divider. The tube potential of X-rays machine MG165 is obtained. The results show the ripple of tube potential is less than 4%, when tube potential is more than 35 kV, the ripple is less than 1%. The relative deviation between average tube potential and nominal value is less than 1.8% in the range from 7.5 kV to 150 kV. The repeatability of average tube potential is better than 0.01%. It is less than 0.02% on the condition of re-open beam and re-open generator. All the testing results shows this X-rays generator is satisfy the requirement for realization, dissemination and international comparison of primary standard of air kerma in the range of low-energy X-rays.
|
|
|
|
|
[1]李锋坦, 李东, 张云亭,等. 管电压对CT值测量、辐射剂量及图像质量影响的模型研究[J]. 中华放射学杂志,2013,47(5):458-461.
[2]JJG 744—2004, 医用诊断X射线辐射源[S].
[3]GB/T 12162.1—2000, 用于校准剂量计和剂量率仪及确定其能量响应的X和γ参考辐射:第1部分:辐射特性及产生方法[S].
[4]吴金杰,王培玮,段小娟,等. 低能X射线基准辐射装置的建立[J]. 计量学报,2011,32(z1):9-13.
[5]段小娟,吴金杰,杨元第,等. EGSnrc程序模拟低能X射线光管参数对射线能谱的影响[J].计量学报,2011,32(z1):51-53.
[6]Burns D T, O’Brien M. Comparison of the NIST and BIPM standards for air kerma in medium-energy X-rays[J]. Journal of Research of the National Institute of Standards & Tec, 2006, 111(5):385-391.
[7]Burns D T, Kessler C, Büermann L. Key comparison BIPM. RI(I)-K2 of the air-kerma standards of the PTB, Germany and the BIPM in low-energy X-rays[J]. Metrologia, 2014, 51(1A):1-10.
[8]Chu C H, Yuan M C, Huang W S, et al. Dosimetry and kVp standardization for quality assurance of mammography[J]. Radiation Physics and Chemistry, 2014, 104(11):316-320.
[9]Vivolo Vitor , Potiens P A, Correa E. Comparesion of kVp and ppv measurements between a constant potential and a clinical x-ray system used for calibration in mammography[C]//2011 International Nuclear Atlantic Conference, MG, Brazil, 2011.
[10]薛永库,杨国山,周红梅,等. 用非介入法检测诊断X射线机的管电压和曝光时间[J].中国辐射卫生,2001,10(4):193-196.
[11]蒋伟,吴金杰,方方,等. 一种便携式X光机的能谱及管电压测量研究[J].核电子学与探测技术,2013,33(10),1229-1233.
[12]Kramer H M, Selbach H J, Iles W J. The practical peak voltage of diagnostic X-ray generators[J]. Br J Radiol, 1998, 71(842):200-209.
[13]Kramer H M, Selbach H J. Extension of the range of definition of the practical peak voltage up to 300 kV [J]. Br J Radiol, 2008, 81:693-698.
[14]杨建, 侯新生, 郑永明,等. 基于非介入式X射线管电压测量的研究[J].中国测试技术, 2008,34(5):110-112.
[15]IEC 61676—2002, International electrotechnical commission. Medical electrical equipment-dosimetric instruments used for non-invasive measurements of X-ray tube voltage in diagnostic radiology[S]. |
|
|
|