Analysis of Failure Modes and the Reasons of Ultrasonic Heat Meter by Durability Test |
DU Wei-peng,CHEN Guo-fu,ZHAO Jian-liang |
Zhejiang Province Institute of Metrology, Hangzhou, Zhejiang 310013, China |
|
|
Abstract Durability test of ultrasonic heat meter (UHM) were carried out, and 5 conditions causing UHM failure were found. Those conditions includes the invalidation of ultrasonic transducer, the measuring tube or ultrasonic reflector fixture made of plastic, calculator, sealing ring of tube, and the temperature sensor. The changes of data showed by UHM are analyzed, under the 5 failure modes mentioned above. The corresponding methods are pointed out to improve the durability of UHM.
|
|
|
|
|
[1]付卫东.智能热量表的技术发展[J].中国仪器仪表,2007, (6):24-28.
[2]李微. 智能超低功耗超声波热量表[J].计量学报, 2013,34(5): 469-473.
[3]赵玉敏,许翠翠,张天亮. 热量表耐久性试验的研究[J]. 计量技术,2013, (7):67-69.
[4]朱江. 热量表流量传感器耐久性试验装置的研制[J]. 计测技术,2014, (5):18-21.
[5]JJG 225—2001 热能表检定规程[S].2001.
[6]CJ 128—2007 热量表[S].2008. |
|
|
|