The Method of Calibrating Rotating Angle of Goniophotometer using Laser Tracker |
CHEN Ting,ZHOU Weng-qing,YE Huai-chu,LU Xin,YE Xin,MAO Zhen-hua |
Zhejiang Institute of Metrology, Hangzhoug, Zhejiang 310018, China |
|
|
Abstract The photometric characteristic of light source is the important parameter in lamp and lantern design and the production processes, while goniophotometer is the key equipment to measure the candela distribution curve. The precision of rotating angle of γ-axis and C-axis in goniophotometer has an important influence on goniophotometry measurement, and those angles cannot be calibrated by the conventional angle measuring instrument because those angles are formed by virtual trajectories in space. The laser tracker is used to measure the rotating angle in space. First, the factors affecting the measurement precision are analyzed, then the measurement parameters of laser tracker are optimized. Later, the 3-D measuring points are transformed to 2-D plane by coordinate conversion. Finally, the angle values are calculated by software and compared with the values read by goniophotometer, so the calibration result is obtained.
|
|
|
|
|
[1]倪凯,牟同升.带旋转反光镜的分布光度测试系统[J].光学仪器,2004,26(3):3-8.
[2]林祥灿.灯具光度分布测量技术分析[J].商品与质量,2012,(11):274-275.
[3]赵伟强,刘慧,刘建,等.高功率LED单管总光通量的测量技术[J].计量技术,2012,(9):25-27.
[4]胡月萍.分布光度计在照明工程设计中的应用[J].光源与照明,2009,(4):25-29.
[5]闫勇刚,欧阳健飞,杨红果,等.激光制导测量机器人自动跟踪算法研究[J].计量学报,2008,29(4A):181-184.
[6]俞建峰,钱建明,顾高浪.LED灯具光通量检测技术的分析研究[J].光学仪器,2011,33(5):9-13.
[7]王建平,郭奉杰.灯具分布光度计现状分析[J].电子测试,2008,(7):46-50.
[8]马骊群,王继虎,曹铁泽,等.6D精密摇摆台的动态参数校准[J].计量学报,2008,29(4A):65-68.
[9]马骊群,王继虎,曹铁泽,等.微波紧缩场面板的精密检测与定位[J].计量学报,2006,27(3A):45-48.
[10]林嘉睿,邾继贵,张皓琳,等.激光跟踪仪测角误差的现场评价[J].仪器仪表学报,2012,33(2):463-468. |
|
|
|