Uncertainty Evaluation for the Determination of Mo in Geological Samples by Spectrophotometr |
ZHANG Ji-guang |
Shenyang Testing and Quality Supervision Center for Geological and Mineral Products, The Ministry of Land and Resources, Shenyang, Liaoning 110032, China |
|
|
Abstract The sources of measurement uncertainty for the analytical results of Mo content in geological samples by spectrophotometry is determined and sub-uncertainties of the analytical results are evaluated through studying the whole process of the analytical results. With the least square method,the calibration curve is fitted. Through the evaluation of the analytical results of various contents of two samples, the sub-uncertainties from the sample repeated measures and the fitting of the calibration curve are the major contributor to the expanded uncertainty. And it shows that the lower the concentrations of Mo, the greater the relative standard uncertainty of analytical results.
|
|
|
|
|
[1]JJF 1059—1999,测量不确定度评定与表示[S].
[2]中国实验室国家认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社,2002.
[3]郑小敏.铋磷钼蓝分光光度法测定铁矿石中磷量的不确定度评定[J].冶金分析,2008,28(1):77-80.
[4]王伟,潘春秀.光度法测定钢铁及合金中锰含量的不确定度评定[J].计量学报,2005,26(2):189-192.
[5]徐夏玲,王雪峰. 二乙基二硫代氨基甲酸银分光光度法测定污水中总砷不确定度评定[J].环境研究与监测,2009,22(2):36-39.
[6]包德君, 魏伯荣, 陈青,等.高效液相色谱测定防老剂D 纯度的不确定度评定[J].计量学报,2007,28(2):174-176.
[7]刘建坤,朱家平,郑荣华.分光光度法测定地下水中亚硝酸根的不确定度评定[J].分析试验室,2009,28(5):113-117.
[8]柯式镇,何亿成,王界益.岩石气体孔隙度测量不确定度分析[J].计量学报,2007,28(2):177-179.
[9]JJG 196—2006,常用玻璃量器检定规程[S].
[10]JJF 1135—2005,化学分析测量不确定度评定[S].
[11]JJG 646—2006,移液器检定规程[S].
[12]GB/T 12807—1991,实验室玻璃仪器分度吸量管[S]. |
|
|
|