New Automatic Apparatus of the Melting Point of Gallium |
SUN Jian-ping1,2,CHEN Wei3,QIU Ping2,LIU Jian-qing3,ZHANG Xin-xin1 |
1. University of Science and Technology Beijing, Beijing 100083, China;
2. National Institute of Metrology, Beijing 100013, China;
3. Hebei University, Baoding, Hebei 071002, China |
|
|
Abstract A new apparatus of the melting point of gallium is developed in National Institute of Metrology.The melting point of gallium can be realized automatically by the fixed point furnace with the three-zone freezing and heating technique.The excellent stability and uniformity of the fixed point furnace is obtained because of the precise temperature control and the reasonable design.The cell of the gallium melting point is filled with the 99.99999% purity gallium that is purified by the present technique and improved perfusion technique.The experiment results show that the plateau of the apparatus lasts more than 70 hours,and the difference of the plateau from 20% to 80% is less than 0.15 mK.The reproducibility of the apparatus is 0.07 mK,and the expanded uncertainty is 0.36 mK(k=2).
|
|
|
|
|
[1]Preston-Thomas H. The International Temperature Scale of 1990(ITS-90)[J]. Metrologia, 1990, 27: 3-10.
[2]孙建平,邱萍,金志军.新一代锌固定点容器的研制[J].计量学报,2007,28(3A):51-54.
[3]Winiatmo J V, Katsuhiko H, Masaru Arai. Development of a temperature fixed point cell using high-purity aluminium[J]. Trans of the Society of Instrument and Control Engineers, 2006, 42(2): 115-122.
[4]Magnum B W, Bloembergen P, Chattle M V, et al. Optimal realization of the defining points of the ITS-90 that are used in contact thermometry[C]//Comité Consultatif de Termométrie,20th Session,Sèvres,France, 2000.
[5]Borovika M, Sostman H E. Melting and triple point temperatures of gallium as fixed points of the international practical temperature scale[J]. Review of Scientific Instruments, 1984, 55(10): 1639-1642.
[6]Chattle M V, Rusby R L, Bonnier G, et al. An intercomparison of sealed cells for realizing the melting and triple points of gallium[J]. Metrologia, 1987, 23(3): 153-159.
[7]Ancsin J. A gallium calibration apparatus[J]. Metrologia 1993,30(2):105-112.
[8]Bojkovski J, Hiti M, Batagelj V, et al. New approach in filling of fixed-point cells: case study of the melting point of gallium[J]. Int J Thermophys,2008,29(1):119-125.
[9]Steur P P M, Dematteis R. Some curious results with a gallium fixed-point cell[J]. Int J Thermophys,2011,32(1-2):285-292.
[10]Steffen R, Martin F. Improved characterization and certification procedure for fixed-point cells[J].计量学报,2008,29(5A):82-85.
[11]邱萍,闫小克,瞿咏梅.镓三相点容器的研制[J].计量学报,2001,22(4):246-248.
[12]闫小克,邱萍,瞿咏梅,等.镓三相点的复现[J].现代计量测试,2001,9(3):34-43. |
|
|
|