Influence of Coriolis Mass Flowmeter Amplitude on Zero Point |
HUANG Ya1,XU Ke-jun1,2,LIU Chen-ci1 |
1. School of Electrical and Automation Engineering, Hefei University of Technology, Hefei, Anhui, 230009, China
2. Engineering Technology Research Center of Industrial Automation, Anhui Province, Hefei, Anhui, 230009, China |
|
|
Abstract Aiming at the problem of the zero drift caused by the change of the Coriolis mass flowmeter amplitude, the flexural vibration equation of the flow tube is first established by simplifying the model, the displacement response of the two detection points is solved, and the linear relationship between the zero phase difference and the amplitude is obtained; Then, the amplified signals of the sensors under different amplitudes are collected, a zero-crossing detection algorithm to calculate the zero-point phase difference is written, a straight line between the zero point and the amplitude is established, and the linear relationship between the zero point and the amplitude through experiments is verified; Finally, combined the verification procedures and the zero point calculation results, the measurement error introduced the zero point drift caused by the amplitude change is analyzed qualitatively and quantitatively, the results show that the greater the amplitude change, the greater the measurement error introduced, when the flow rate is 10kg/min, the amplitude change 10mV introduces a measurement error of 0.727%.
|
Received: 06 December 2021
Published: 21 February 2023
|
|
|
|
|
[8] |
张启平, 魏祥毓, 唐秋林, 等. 科里奥利质量流量计的激振与信号提取 [J]. 仪器仪表与分析监测, 1992 (2): 1-4.
|
[10] |
刘翠, 侯其立, 熊文军, 等. 面向微弯型科氏质量流量计的高精度过零检测算法实现 [J]. 电子测量与仪器学报, 2014, 28 (6): 675-682.
|
[4] |
Wang L, Hu L, Fu X, et al. Experimental investigation on zero drift effect in Coriolis mass flowmeters[C]// FLOMEKO, Taipei, China, 2010.
|
[7] |
刘文, 徐科军, 方正余, 等. 用于振幅控制的科氏流量管模型的实验识别 [J]. 电子测量与仪器学报, 2018 (6): 39-45.
|
|
Ren J X, Qian X L, Qu Y Q, et al. Research on the influence of Coriolis mass flowmeter vibrator balance characteristics on zero drift [J]. Journal of Transducer Technology, 2016, 29 (9): 1311-1316.
|
[3] |
Enz S, Thomsen J J, Neumeyer S. Experimental investigation of zero phase shift effects for Coriolis flowmeters due to pipe imperfections [J]. Flow Measurement and Instrumentation, 2011, 22 (1): 1-9.
|
[5] |
Yaushev A A, Taranenko P A, Loginovskiy V A. Analysis of influence of mechanical boundary conditions on zero point shift of Coriolis flowmeter [J]. Procedia Engineering, 2017, 206: 552-557.
|
|
Hou Q L, Xu K J, Li Y, et al. Development of a high-precision Coriolis mass flow transmitter based on TMS320F28335 [J]. Chinese Journal of Scientific Instrument, 2010 (12): 2788-2795.
|
[11] |
黄雅, 徐科军, 刘陈慈, 等. 基于特性图分析的科氏质量流量计模拟驱动研究[J]. 计量学报, 2021, 42(2): 189-198.
|
[12] |
JJG 1038-2008科里奥利质量流量计检定规程[S]. 2008.
|
|
Ma Y P, Shi C, Luo Y, et al. Numerical simulation of gas-liquid two-phase flow in Coriolis flowmeter [J]. Acta Metrology, 2018, 39 (5): 663-667.
|
[16] |
马也驰, 赵伟国, 章圣意. 基于回波信号相似度的气体超声流量计动态阈值法研究 [J]. 计量学报, 2022, 43(4): 482-488.
|
[2] |
任建新, 钱小亮, 屈雅琴, 等. 科氏质量流量计振子平衡特性对零点漂移的影响研究 [J]. 传感技术学报, 2016, 29 (9): 1311-1316.
|
[9] |
侯其立, 徐科军, 李叶, 等. 基于 TMS320F28335 的高精度科氏质量流量变送器研制 [J]. 仪器仪表学报, 2010 (12): 2788-2795.
|
|
Xu H R, Xu K J, Zhang L, et al. Research on Key Technology of Coriolis Mass Flowmeter for Measuring Flow Rate of Gas Containing Liquid [J]. Acta Metrology Sinica, 2021, 42 (4): 483-494.
|
[17] |
Krishna R K, Unnikrishnan M, Kochupillai J. A finite element model for the analysis of flexible tube Coriolis mass flow meter using first order shear deformation shell theory [J]. Flow Measurement and Instrumentation, 2021, 79: 101947.
|
[1] |
Kolahi K, Storm R, Rck H. Detection of zero shift in Coriolis Mass Flowmeters[C]//Proceedings of the XVI IMEKO World Congress, 2000, 6: 61-66.
|
[6] |
刘延柱, 陈文良, 陈立群.振动力学[M].北京: 高等教育出版社,1998: 201-213.
|
|
Liu W, Xu K J, Fang Z Y, et al. Experimental identification of Coriolis flow tube model for amplitude control [J]. Journal of Electronic Measurement and Instrument, 2018 (6): 39-45.
|
|
Zhang Q P, Wei X Y, Tang Q L, et al. Vibration and signal extraction of Coriolis mass flowmeter [J]. Instrumentation, Analysis and Monitoring, 1992 (2): 1-4.
|
|
Huang Y, Xu K J, Liu C C, et al. Research on Coriolis Mass Flowmeter Simulation Drive Based on Characteristic Diagram Analysis[J]. Acta Metrology Sinica, 2021, 42(2): 189-198.
|
[13] |
孙立军, 党士忠, 张涛, 等. 流量计性能受流量稳定性影响实验 [J]. 天津大学学报: 自然科学与工程技术版, 2018, 51 (4): 422-432.
|
[14] |
徐浩然,徐科军,张伦, 等. 科氏质量流量计测量含气液体流量关键技术综述[J]. 计量学报, 2021, 42 (4): 483-494.
|
|
Liu C, Hou Q L, Xiong W J, et al. Implementation of high-precision zero-crossing detection algorithm for micro-bending Coriolis mass flowmeters [J]. Journal of Electronic Measurement and Instrument, 2014, 28 (6): 675-682.
|
[15] |
Krishna R K, Unnikrishnan M, Kochupillai J. A finite element model for the analysis of flexible tube Coriolis mass flow meter using first order shear deformation shell theory [J]. Flow Measurement and Instrumentation, 2021, 79: 101947.
|
|
Ma Y C, Zhao W G, Zhang S Y. Study on Dynamic Threshold Method Based on Echo Similarity for Ultrasonic Gas Flow Meter [J]. Acta Metrologica Sinica, 2022, 43 (4): 482-488.
|
|
Sun L J, Dang S Z, Zhang T, et al. Experiment of flowmeter performance affected by flow stability [J]. Journal of Tianjin University: Natural Science and Engineering Technology Edition, 2018, 51 (4): 422-432.
|
[14] |
马亚朋, 石础, 罗宇, 等. 科氏流量计气-液两相流的数值模拟 [J]. 计量学报, 2018, 39 (5): 663-667.
|
|
|
|