Research on the Factors Influencing the Calibration of Medical Diagnostic Half-value Layer Instrument
WEI Peng1,WANG Jia-wei2,LI De-hong3,MA Bo-xuan3,GUO Bin3
1.Fujian Institute of Metrology,Fuzhou,Fujian 350003,China
2.Anhui Institute of Metrology,Hefei, Anhui 230051,China
3.National Institute of Metrology,Beijing 100029,China
Download: PDF (558 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  The half-value layer measurement is required in various diagnostic X-ray radiation source verification regulations or calibration specifications, and half value layer measuring instruments are also widely used, but the calibration method of half value layer measuring instruments has not been clear. The main influencing factors in the calibration of the half value layer measuring instrument are discussed. The results show that the detector angle has the greatest influence on the measurement results. Therefore, the half value layer measuring instrument to be calibrated should be placed in the center of the test plane during calibration, and the angle deviation should not exceed ±1°.
Key wordsmetrology;HVL;RQR;diagnostic X-ray radiation source      calibration     
Received: 08 February 2021      Published: 13 January 2023
PACS:  TB98  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
WEI Peng
WANG Jia-wei
LI De-hong
MA Bo-xuan
GUO Bin
Cite this article:   
WEI Peng,WANG Jia-wei,LI De-hong, et al. Research on the Factors Influencing the Calibration of Medical Diagnostic Half-value Layer Instrument[J]. Acta Metrologica Sinica, 2023, 44(1): 127-131.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2023.01.18     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2023/V44/I1/127
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech