Primary Vibration Calibration of High Temperature Piezoelectric Acceleration Sensor |
WANG Jia-hao1,ZUO Ai-bin2,PENG Yue-xiang1 |
1. Beijing University of Technology, Beijing 100022, China
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract For the problem that the sensitivity of the piezoelectric acceleration sensor increases with the increase of temperature in a high temperature environment,a sine-approximation absolute calibration method for the acceleration sensor is proposed and a related experimental system is built.The sensitivity measurement and calibration test of the high temperature resistant acceleration sensor are carried out by the mentioned method under different temperature environments from room temperature to 400℃.The experimental results show that the sensitivity of the acceleration sensor shows a linear increase trend when the ambient temperature rises.
|
Received: 27 August 2021
Published: 28 December 2022
|
|
|
|
|
[1]高金芳. 振动冲击传感器的温度响应校准 [J]. 宇航计测技术, 1993, 12 (2): 58-66.
Gao J F. Temperature response calibration ofvibration shock sensor [J]. Journal of Astronautic Metrology and Measurement, 1993, 12 (2): 58-66.
[2]高金芳. 振动传感器温度响应的激光绝对校准 [J].计量学报, 1994, 15 (1): 1-7.
Gao J F. Laser Absolute Calibration of Vibration Sensor Temperature Response [J]. Acta Metrologica Sinica, 1994, 15 (1): 1-7.
[3]杜晓辉, 刘帅, 朱敏杰, 等. 高温压电式加速度传感器的温度响应测试评价 [J]. 压电与声光, 2020, 42 (3):381-385.
Du X H, Liu S, Zhu M J, et al. Temperature response test evaluation of high-temperature piezoelectric acceleration sensor [J]. Piezoelectrics & Acoustooptics, 2020, 42 (3):381-385.
[4]于梅, 孙桥, 冯源, 等. 正弦逼近法振动传感器幅相特性测量技术的研究 [J]. 计量学报, 2004, 25 (4): 344-348.
Yu M, Sun Q, Feng Y, et al. A study of application of the sine-approximation method tothe primary vibration calibration [J]. Acta Metrologica Sinica, 2004, 25 (4):344-348.
[5]ISO 16063-34: Methods for the calibration of vibration and shock transducers—Part34: Testing ofsensitivity at fixed temperatures[S]. 2019.
[6]刘泽鹏, 左爱斌, 彭月祥, 等.基于振动传感器温度灵敏度校准的温控系统研究 [J]. 计量技术, 2016(10)21-23.
[7]刘杰坤, 马修水, 马勰. 激光多普勒测振仪研究综述 [J]. 激光杂志, 2014, 35 (12): 1-5.
Liu J K, Ma X S, et al. Review of laser Doppler vibrometer [J]. Laser Journal, 2014, 35 (12): 1-5.
[8]杨明, 蔡晨光, 王颖, 等.外差式激光干涉信号的采集与解调方法综述[J]. 激光杂志, 2018, 39 (1): 20-24.
Yang M, Cai C G, Wang Y, et al. Overview of the collection and demodulation methods of heterodyne laser interference signals [J]. Laser Journal, 2018, 39 (1): 20-24.
[9]梁志国. 一种基于残周期拟合的激光测冲击信号解调方法[J]. 计量学报, 2021, 42(4): 503-510.
Liang Z G. A Novel Demodulation Method Based on Partial Period Curve-fitting for FM Signal in Impulse Acceleration Calibration by Using Laser Interference Methods[J]. Acta Metrologica Sinica, 2021, 42(4): 503-510.
[10]Sun Q, Bruns T, Tubner A, et al. Modifications of the sine-approximation method for primary vibration calibration by heterodyne interfero-metry [J]. Metrologia, 2009, 46 (6): 646-654.
[11]杨明,蔡晨光,刘志华, 等. 基于外差激光干涉法的三轴向振动绝对校准方法研究[J]. 计量学报, 2018, 39(2): 201-206.
Yang M, Cai C G, Liu Z H, et al. Research on the Method of the Triaxial Primary Vibration Calibration Using the Heterodyne Interferometry[J]. Acta Metrologica Sinica, 2018, 39(2): 201-206.
[12]王朝, 蔡晨光, 杨明, 等. 基于激光干涉仪数字信号解码的振动校准方法 [J]. 计量学报, 2021, 42(3): 282-286.
Wang Z, Cai C G,Yang M, et al. Vibration Calibration Method Based on the Decoding of Heterodyne Interferometer Output S/PDIF Digital Signal [J]. Acta Metrologica Sinica, 2021, 42 (3): 282-286.
[13]ISO 16063-11: Methods for the calibration of vibration and shock transducers—Part11: Primary vibration calibration by laser interferometr[S]. 1999. |
|
|
|