Calibration of Low Frequency Accelerometer Based on the Earth‘s Graviation Method |
YANG Ming1,ZHAO Huan-xing2,LIU Zhi-hua3,LI He-jia4,CAI Chen-guang3 |
1. The Electrical Engineering College, Guizhou University, Guiyang, Guizhou 550025, China
2. Unit 32382 of PLA, Wuhan, Hubei 432200, China
3. National Institute of Metrology, Beijing 100029, China
4. Inner Mongolia Institute of Metrology Testing and Research, Hohhot, Inner Mongolia 010050, China |
|
|
Abstract The earth gravity method described in ISO 16063-16 can only be used for static calibration of low-frequency accelerometer sensitivity amplitude, and cannot be used for dynamic calibration of sensitivity amplitude and phase. The machine vision method is adopted to measure the phase of the excitation acceleration and determine that of the output signal of the accelerometer at the corresponding time, and the dynamic calibration of sensitivity magnitude and phase based on earth’s graviation method is accomplished. The rotator provides the excitation acceleration with a constant peak 1g for the accelerometer, which efficiently avoids the influence of the output low SNR of the accelerometer caused by the limited excitation acceleration on the calibration accuracy. Experimental results show the investigated earth’s graviation method can achieve the high-accuracy sensitivity magnitude and phase calibration in the range from DC to 10Hz.
|
Received: 05 February 2021
Published: 20 August 2022
|
|
|
|
|
[1]熊春宝, 路华丽, 朱劲松, 等. 基于GPS RTK和加速度计的桥梁动态变形监测试验[J]. 振动与冲击, 2019, 38(12): 69-73.
Xiong C B, Lu H L, Zhu J S, et al. Dynamic deformation monitoring of bridge structures based on GPS-RTK and accelerometers [J]. Journal of Vibration and Shock, 2019, 38(12): 69-73.
[2]李宏男, 田亮, 伊廷华, 等. 大跨斜拱桥结构健康监测系统的设计与开发[J]. 振动工程学报, 2015, 28(4): 574-584.
Li H N, Tian L, Yi T H, et al. Design and development of structural health monitoring system for long span skew arch bridge [J]. Journal of Vibration Engineering, 2015, 28(4): 574-584.
[3]Sabato A, Niezrecki C, Fortino G. Wireless MEMS-Based Accelerometer Sensor Boards for Structural Vibration Monitoring: A Review[J]. IEEE Sensors Journal, 2017, 17(2): 226-235.
[4]汤宝平, 罗雷, 邓蕾, 等. 风电机组传动系统振动监测研究进展[J]. 振动、测试与诊断, 2017, 37(3): 417-425.
Tang B P, Luo L, Deng L, et al. Research progress on vibration monitoring of wind turbine transmission system [J]. Journal of Vibration, Measurement & Diagnosis, 2017, 37(3): 417-425.
[5]ISO 16063-16. Methods for the calibration of vibration and shock transducers-part 16: Calibration by Earth’s gravitation [S]. 2011.
[6]于梅, 刘爱东, 马明德, 等. 超低频激光干涉法振动幅相特性测量技术的研究[J]. 振动与冲击, 2011, 30(11): 130-134.
Yu M, Liu A D, Ma M D, et al. Magnitude and phase measurement technology for ultra-low frequency vibration using laser interferometry [J]. Journal of Vibration and Shock, 2010, 29(4): 107-110.
[7]鲁敏, 胡红波. 基于正弦激励的压电加速度计模型参数辨识[J]. 计量学报, 2020, 41(1): 55-59.
Lu M, Hu H B. Calibration of Piezoelectric Accelerometer’s Calibration Dynamic Characteristics using Primary Vibration Calibration [J]. Acta Metrologica Sinica, 2020, 41(1): 55-59.
[8]Scott D A, Dickinson L P. Distortion effects in primary calibration of low-frequency accelerometers [J]. Metrologia, 2014, 51: 212-224.
[9]Garg N, Schiefer M I. Low frequency Accelerometer Calibration using an optical encoder sensor [J]. Measurement, 2017, 111: 226-233.
[10]Bruns T, Gazioch S. Correction of shaker flatness deviations in very low frequency primary accelerometer calibration [J]. Metrologia, 2016, 53: 986-990.
[11]Yang M, Wang Y, Cai C G, et al. Monocular vision-based low-frequency vibration calibration method with correction of the guideway bending in a long-stroke shaker [J]. Optics Express, 2019, 27(11): 15968-15981.
[12]董春梅, 陈希军, 刘庆博, 等. 正交双加速度计两种安装位置在重力场中的标定方法 [J]. 中国惯性技术学报, 2014, 22(5): 693-700.
Dong C M, Chen X J, Liu Q B, et al. Calibration method for dual orthogonal accelerometers at two combinational installing positions in 1g field [J]. Journal of Chinese Intertial Technology, 2014, 22(5): 693-700.
[13]占伟伟, 蔡莉. 一种减小加速度计重力法标定中安装误差影响的方法 [J]. 振动与冲击, 2019, 38(13): 249-252.
Zhan W W, Cai L. A method reducing effect of installation error in accelerometer gravimetric calibration [J]. Journal of Vibration and Shock, 2019, 38(13): 249-252.
[14]Dosch J. Low Frequency Accelerometer Calibration Using Earth’s Gravity [J]. IMAC XXV: Conference & Exposition on Structural Dynamics, 2007, 1-7.
[15]Yang M, Zhu H J, Cai C G, et al. Monocular vision-based Earth’s graviation method used for low-frequency vibration calibration [J]. IEEE Access, 2020, 8: 129087-129093.
[16]Li W M, Jin J, Li X F, et al. Method of rotation angle measurement in machine vision based on calibration pattern with spot array [J]. Applied Optics, 2010, 49(6): 1001-1006.
[17]Dong H X, Fu Q, Zhao X, et al. Practical rotation angle measurement method by monocular vision [J]. Applied Optics, 2015, 54(3): 425-435. |
|
|
|