Acta Metrologica Sinica  2022, Vol. 43 Issue (5): 629-635    DOI: 10.3969/j.issn.1000-1158.2022.05.11
Current Issue | Archive | Adv Search |
Study on the Determination of Apparent Density ConversionStandard Density of Crude Oil at 15~80℃ by Comparing Oscillating Tube and Pycnometer Method
HE Yun-teng,ZHANG Jian,LI Hua,LI Dong-hui,WU Ying-xiang
Institute of Mechanics, Chinese Academy of Sciences, Beijing 100190, China
Download: PDF (3139 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  Crude oil standard density (20℃) is an important item in the physical and chemical indicators of crude oil. By using Anton Paar (Anton Paar) DMATM 35 V4 Ex Petrol explosion-proof portable density meter and other equipment, the density of crude oil samples from oil wells in multiple oilfields were detected by pyknometer method and vibrating-tube densitometer method respectively at 15~80 ℃, the density temperature curve was drawn, the empirical formula was deduced and improved, and a calculation formula with test density and temperature as parameters was given. The oscillating tube density test method and the petroleum meter were combined to calculate the crude oil standard density, and the relative error was controlled within ±0.5%. It is possible to quickly obtain crude oil standard density parameters on-site.
Key wordsmetrology      crude oil standard density      vibrating-tube density meter      explosion-proof portable      on-site inspection      calculation method     
Received: 17 August 2020      Published: 18 May 2022
PACS:  TB933  
Fund:Strategic Priority Research Program of Chinese Academy of Sciences(type B)
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
HE Yun-teng
ZHANG Jian
LI Hua
LI Dong-hui
WU Ying-xiang
Cite this article:   
HE Yun-teng,ZHANG Jian,LI Hua, et al. Study on the Determination of Apparent Density ConversionStandard Density of Crude Oil at 15~80℃ by Comparing Oscillating Tube and Pycnometer Method[J]. Acta Metrologica Sinica, 2022, 43(5): 629-635.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2022.05.11     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2022/V43/I5/629
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech