Comparison of Scale Division Methods between Single-fixed-point Extrapolation and Multi-fixed-point Interpolation for 0.9μm Pyrometers
LIU Xiao-ying1,2,LIU Ju-fen2,WANG Jing-hui1,LU Xiao-feng1
1. Heat Division, National Institute of Metrology, Beijing 100029, China
2. School of Information and Communication Engineering, North University of China, Taiyuan, Shanxi 030051, China
Download: PDF (1244 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  Scale division methods for single fixed-point extrapolation and multiple fixed-point interpolation are two commonly used calibration methods for standard photoelectric pyrometers. A comparative study of two methods was carried out for a newly developed 0.9μm photoelectric pyrometer. In the temperature measurement range of 500~1800℃, the high-temperature blackbody furnace was used as a comparison source to verify the value at the hundreds of temperature points against a standard photoelectric pyrometer. The results show that the consistency of the temperature values of the two methods is better than 0.2℃, and the maximum error away the standard value is not more than 0.6℃. The calibration uncertainty of the two methods is evaluated as 0.2~1.3℃ and 0.2~0.8℃, respectively, k=2. The single fixed point extrapolation method must measure the pyrometer spectral responsivity and non-linearity. The multiple fixed-point interpolation method is simple and easy to operate, and the uncertainty can be kept at a high level. The multi-fixed-point interpolation method can be universally used for most temperature calibration laboratories, whether it is for the manufacture or the periodic calibration.
Key wordsmetrology      scale division method      photoelectric pyrometer      single-fixed-point extrapolation      multi-fixed-point interpolation      high temperature fixed point      uncertainty     
Received: 13 April 2020      Published: 20 August 2021
PACS:  TB942  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
LIU Xiao-ying
LIU Ju-fen
WANG Jing-hui
LU Xiao-feng
Cite this article:   
LIU Xiao-ying,LIU Ju-fen,WANG Jing-hui, et al. Comparison of Scale Division Methods between Single-fixed-point Extrapolation and Multi-fixed-point Interpolation for 0.9μm Pyrometers[J]. Acta Metrologica Sinica, 2021, 42(8): 1000-1005.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2021.08.04     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2021/V42/I8/1000
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech