The Influence of Air Inlet Flow Rate on the Capture Efficiency of PM2.5 Cutter |
LIU Jia-qi,ZHANG Guo-cheng,ZHAO Xiao-ning,WU Dan,JING Wen-jie |
Beijing Institute of Metrology, Beijing 100029, China |
|
|
Abstract Generally, the commercial cutters have a fixed working flow rate range, if the actual gas flow rate into the cutter is not in this fixed range, deviation of the capture efficiency might be occurred. In this study, by using the cutter capture efficiency evaluation system based on static box method, the capture efficiency for two domestic and imported PM2.5 cutters were evaluated. The influences of the standard working flow of the cutter and the flow of 70%, 80%, 120% and 130% on the trapping efficiency of the PM2.5 cutter were studied. The results showed that with the change of inlet flow rate, the change trend of Da50 is opposite. The distribution range of Da50 is 140% to 80% of the value under standard operating flow. The geometrical standard deviations of the penetrating curve staynearly unchanged, while the change is not more than 0.9% compared with the value under standard operating flow. This study has scientific value for further regulating the manufacture and the usage of cutter.
|
Received: 19 May 2020
Published: 20 April 2021
|
|
|
|
|
[1]张国城, 沈正生, 赵红达, 等. 粉尘性质对光散射式粉尘仪测量结果的影响 [J]. 计量技术, 2017, (7): 46-48.
Zhang G C, Shen Z S, Zhao H D, et al. The influence of dust properties on the measurement results of light scattering dust meter [J]. Measurement Technique, 2017, (7): 46-48.
[2]Lin C W, Chen T J, Huang S H, et al. Effect of Aerosol Loading on Separation Performance of PM2.5 Cyclone Separators [J]. Aerosol and Air Quality Research, 2018, 18(6): 1366-1374.
[3]Fischer J. Progress towards a new definition of the kelvin [J]. Metrologia, 2015, 52 (5): 364-375.
[4]张国城, 赵晓宁, 靳钰. 两类PM2.5监测仪及其计量检测方法的差异 [J]. 计量技术, 2016, (8): 39-42.
Zhang G C, Zhao X N, Jin Y. Differences between two types of PM2.5 monitors and the measurement methods [J]. Measurement Technique, 2016, (8): 39-42.
[5]刘彬,李德健,赵志彪,等. 改进的快速跟踪回声状态网络及PM2.5预测 [J]. 计量学报, 2020, 41(9): 1138-1145.
Liu B, Li D J, Zhao Z B, et al. Improved Fast Tracking Echo State Network and PM2.5 Prediction[J]. Acta Metrologica Sinica, 2020, 41(9): 1138-1145.
[6]US EPA. EPA-HQ-OAR-2004-0018. 40 CFR Parts 53 and 58, Revised Requirements for Designation of Reference and Equivalent Methods for PM2. 5 and Ambient Air Quality Surveillance for Particulate Matter [S]. 2015.
[7]阮兵,李兴华,谢岩,等. PM2. 5旋风切割器的性能测试与模拟[J]. 环境科学学报, 2018, 38(7): 2811-2817.
Ruan B, Li X H, Xie Y, et al. Performance testing and modeling of PM2. 5 cyclones[J]. Acta Scientiae Circumstantiae, 2018, 38(7): 2811-2817.
[8]陈仲辉, 张文阁, 黄星亮, 等. PM2.5切割器切割特性研究 [J]. 中国计量, 2014, (8): 82-85.
Chen Z H, Zhang W G, Huang X L, et al. Study on cutting characteristics of PM2.5 cutter [J]. China Metrology, 2014, (8): 82-85.
[9]刘泽华, 聂泽东, 刘攀超, 等. 基于数值模拟仿真的PM2.5切割器校准装置的研制[J]. 检验检疫学刊, 2017, (6): 67-70.
Liu Z H, Nie Z D, Liu P C, et al. Development of Calibration Device of PM2. 5 Separator [J]. Journal of Inspection and Quarantine, 2017, (6): 67-70.
[10]李斌, 梁珺成, 杨志杰, 等. 一种新型人工放射性气溶胶发生装置的研制[J]. 计量学报, 2016, 37(z1): 120-123.
Li B, Liang J C, Yang Z J, et al. Development of a New Artificial Radioactive Aerosol Generator [J]. Acta Metrologica Sinica, 2016, 37(z1): 120-123.
[11]吴付祥, 陈建阁, 王静宇, 等. 采样器气体流量闭环控制研究 [J]. 计量学报, 2017, 38(1): 87-90.
Wu F X, Chen J G, Wang J Y, et al. Development of Dust Sampler Based Flow Closed-loop Control [J]. ActaMetrologicaSinica, 2017, 38(1): 87-90.
[12]HJ 93-2013环境空气颗粒物(PM10和PM2. 5)采样器技术要求及检测方法 [S]. 2013.
[13]许金鑫, 由强. 任意阶次多项式最小二乘拟合不确定度计算方法与最佳拟合阶次分析 [J]. 计量学报, 2020, 41(3): 388-392.
Xu J X, You Q. Uncertainty Calculation for Arbitrary Order Polynomial Least-square Fitting and Analysis of the Best Fitting Order [J]. Acta Metrologica Sinica, 2020, 41(3): 388-392.
[14]李新颖, 刘凯, 黄海燕. 多线函数法在曲线拟合中的应用研究[J]. 计量学报, 2018, 39(5): 716-719.
Li X Y, Liu K, Huang H Y. Application of Multi-line Function Method in Curve Fitting [J]. Acta Metrologica Sinica, 2018, 39(5): 716-719.
[15]张文阁, 刘巍, 许萧, 等. PM2. 5监测仪检测用国家一级标准物质的研制 [J]. 计量学报, 2019, 40(1): 159-163.
Zhang W G,Liu W,Xu X,et al. Preparation of National Primary Reference Materials for PM2.5 Monitors [J]. Acta Metrologica Sinica, 2019, 40(1): 159-163.
[16]DUP, LIU J G, GUO H Q, et al. Development of a static test apparatus for evaluating the performance of three PM2.5 separators commonly used in China [J]. Journal of Environmental Sciences, 2020, 32(1): 238-249.
[17]Kuo F Y, Lin Y C, Ke L Y, et al. Detection of Particulate Matter of Size 2. 5 μm with a Surface-Acoustic-Wave Sensor Combined with a Cyclone Separator [J]. Micromachines, 2018, 9(8): 398-408.
[18]Ewan C, Matthew D B, Michael S, et al. Development and characterization of a high-efficiency, aircraft-based axial cyclone cloud water collector[J]. Atmospheric Measurement Technique, 2018, 11(9): 5025-5048. |
|
|
|