Abstract  According to the research status and application requirements for energy measurement, the reference calorimeter of natural gas is studied.The calorific value of high-purity methane is determined by the oxygen bomb calorimeter, an adiabatic oxygen bomb calorimeter calibrated by the CRM of benzoic, which calorific value of the benzoic CRM traces to SI unit via the national standard.The device sample collect and equilibrium designed by NIM is described.The superior calorific value of methane is 39 900 J/m3 with an expanded rative uncertainty of (the expanded factor,0.6% k=2).
Key wordsMetrology      Energy measurement      Natural gas      Methane      Benzoic      Reference material      Calorific value     
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
LI Jia
SUN Guo-hua
WANG Hai-feng
MENG Fan-min
Cite this article:   
LI Jia,SUN Guo-hua,WANG Hai-feng, et al. Study on the Reference Calorimeter and Method for Determining Calorific Value of Natural Gas Based on Oxygen Bomb Calorimeter[J]. Acta Metrologica Sinica, 2013, 34(6): 592-596.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2013.06.17     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2013/V34/I6/592
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech