A Calibration Method for Inner Installed On-line Monitoring Instrument of Pure Water Machine |
ZHANG Yao-qi,REN Yun,LI Ya,HE Jun-bin |
Hunan Institute of Metrology and Test, Changsha, Hunan 410014, China |
|
|
Abstract To meet with the needs of on-site calibration of monitoring instruments installed in pure water machines, a measurement standard system including high-precision conductivity meter, total organic carbon analyzer and adjustable flow buffer device has been developed.By comparing the indication values of built-in monitoring instruments of pure water machines with the measurement results of the measurement standard system, the on-site calibration of conductivity or resistivity meter and total organic carbon analyzer installed in pure water machines was achieved.The experimental results of eleven pure water machines showed the indication errors of built-in conductivity or resistivity meters ranging from -2.4% to 15% with the expanded uncertainty of 1.7% (k=2) and those of built-in total organic carbon analyzer ranging from -98% to 125% with the expanded uncertainty of 4.8% (k=2).
|
Received: 21 June 2019
Published: 19 December 2019
|
|
|
|
|
[1]GB/T 33087—2016仪器分析用高纯水规格及试验方法[S].
[2]Petra S, Bruno R, Yves G, et al.New approach to calibrating conductivity meters in the low conductivity range[J]. Accreditation & Quality Assurance, 2005, 10(3): 78-81.
[3]Song X P, Spitzer P, Sudmeier U.Novel method for bulk resistance evaluation in conductivity measurement for high-purity water[J]. Accreditation & Quality Assurance, 2007, 12(7): 351-355.
[4]Zhang B, Lin Z, Zhang X. Absolute measurement system for low electrolytic conductivity of aqueous solutions based on van der Pauws theory using a flow-through cell[J]. Accreditation & Quality Assurance, 2014, 19(6): 433-438.
[5]晏天, 梁杰, 赵东蕾, 等. 温盐深测量仪电导率传感器校准的优化与提升[J]. 中国测试, 2018, 44(12): 51-54.
Yan T, Liang J, Zhao D L, et al.Optimisation and improvement of calibration conductivity sensor of the conductivity temperature and depth measuring instruments[J]. China Measurement & Test, 2018, 44(12): 51-54.
[6]宋小平. 溶液电导率的绝对测量方法[J]. 化学分析计量, 2004, 13(6): 88-89.
Song X P.Absolute messurement methods for electrolytic conductivity of solution[J]. Chemical Analysis and Meterage, 2004,13(6): 88-89.
[7]Elena O, Francesca D, Massimo O. Development of an in-line calibration system for flow-through cells for low electrolytic conductivity values[J]. Accreditation & Quality Assurance, 2014, 19(1): 11-16.
[8]魏佳莉. 基于Van Der Pauw法的水溶液电导率绝对测量系统研究[D]. 杭州: 浙江大学, 2013.
[9]林桢, 张潇, 魏佳莉, 等. 基于范德堡法的溶液电导率绝对测量方法[J]. 计量学报, 2015, 36(2): 176-180.
Lin Z, Zhang X, Wei J L, et al.Absolute measurement method for Electrolytic Conductivity solution based on Van Pauw theory[J]. Acta Metrologica Sinica, 2015, 36(2): 176-180.
[10]于得水. 纯水机校准方法初探[J]. 西安文理学院学报(自然科学版), 2017, 20(4): 66-69.
Yu D S. Discussion on the Calibration Method for Water Purification Machine[J]. Journal of Xian University (Natural Science Edition), 2017, 20( 4): 66-69.
[11]Freek B, Niels E D, Eva D. Primary methods for the measurement of electrolytic conductivity[J]. Accreditation & Quality Assurance, 2003, 8(7-8): 346-353.
[12]邱灵佳, 黄国林, 苏玉, 等. 总有机碳测定方法研究进展[J]. 广东化工, 2015, 42(9): 107-108.
Qiu L J, Huang G L, S U Y, et al. An Overview on Current Measurment Methods of Total Organic Carbon[J]. Guangdong chemical industry, 2015, 42(9): 107-108.
[13]王二福.温度对超纯水电导率测量的影响及其对策[J]. 华北电力技术, 1994,(06): 5-7.
Wang E F.The effect and countermeasures of temperature on the electrolytic conductivity measurement of pure water[J]. North China Electric Power, 1994, (06): 5-7.
[14]JJF(湘)05—2018纯水/超纯水系统监测仪表(电导率)计量校准规范[S].
[15]严湘青, 吴旭梅.薄膜-电导率法总有机碳分析仪测量结果不确定度的评定[J]. 药物分析杂志, 2008, 28(12): 2140-2142.
Yan X Q, Wu X M. Evaluation of measurement uncertainty for total organic carbon analyzer with membrane conductmetric detection techonology[J]. Chin J Pharm Anal, 2008, 28(12): 2140-2142.
[16]JJG 821—2005 总有机碳分析仪[S]. |
|
|
|