Design and Development of the WR-15 (50~75) GHz Millimeter-wave Power National Primary Standard |
CUI Xiao-hai1, LI Yong1, ZHANG Shu-yi2, SUN Wei-jie2 |
1. National Institute of Metrology, Beijing 100029, China;
2. Beijing Jiaotong Univercity, Beijing 100044, China |
|
|
Abstract The design and development technique of a WR-15 microcalorimeter based on a thermistor mount is presented. The microcalorimeter is designed as twn-line structure which can lower the influnece from the laboratory temperature floating. The basic principle of the measurement is presented too. The microcalorimeter will be national primary standard of WR-15 (50~75 GHz) millimeter-wave power. It enhanced China RF power measurement capability up to 75 GHz.
|
|
|
|
|
[1] Crowley T P, Cui X H.Design and evaluation of a WR-15(50 to 75 GHz) microcalorimeter[C]// CPEM 2008 Conference,Broomfield,Colorado,USA, 2008.
[2] Allen J W, Fred R C, Larsen N T, et al. The NIST Microwave Power Standards in Waveguide[R]. NIST Technical Note 1511, 1999.
[3] Cui X H, Crowley T P.Comparison of Experimental Techniques for Evaluating the Correction Factor of a Rectangular Waveguide Microcalorimeter[C]//IEEE Transactions on Instrumentation and Measurement,Daejeon, Korea,2010. |
|
|
|