Research on Kα Energy Spectrum and Fluorescence Yield of Samarium |
HUO Bin-bin1,2,CHEN Cheng2,3,WU Jin-jie2,ZUO Guo-ping1,ZHANG Shuai2,4,LI Ting2,3 |
1. University of South China, Hengyang, Hunan 421000, China
2. National Institute of Metrology, Beijing 100029, China
3. Chengdu University of Technology, Chengdu, Sichuan 610059, China
4. China University of Petroleum, Beijing 102200, China |
|
|
Abstract The relationship between fluorescence quality and thickness of radiator Sm2O3 or secondary filter CeO2 by simulating fluorescence instrument with MCNP program was described. When the thickness of Sm2O3 reached 0.4mm, K fluorescence yield is saturated. When the thickness of CeO2 was 0.55mm, Kα fluorescence purity optimum value is 96.6%. The results show that the higher X-ray tube voltage is, the bigger fluorescence yield is. Fluorescence purity reached optimum 82.26% when X-ray tube voltage was 80kV.
|
Received: 23 May 2016
Published: 27 September 2017
|
|
|
|
|
[1]陈成, 吴金杰, 周四春, 等. K 荧光能谱及荧光产额 MC 模拟[J]. 核电子学与探测技术, 2015,35(3):299-302.
[2]Pasqualini G, De Chiara P, Frontera F. Development status of Bragg hard—X-ray concentrators for space astronomy[J]. Ⅱ Nuovo Cimento C,1992,15(5):879-887.
[3]冼鼎昌. 同步辐射的现状和发展[J]. 中国科学基金, 2005, (6):321-325.
[4]于智清, 李向春, 吴广林. 新型透射阳极K荧光X射线装置替代医用放射性辐射源的研究和应用[J]. 辽宁中医学院学报, 2005,7(3):211.
[5]华北辐射防护研究所二室五组. 6至100千电子伏仪器校准用低能光子源—宽束K荧光X射线源[J]. 辐射防护,1979, (5):49-54.
[6]吴金杰,郭彬,李兵,等. 介入法测量工业X射线光机管电压的研究 [J]. 计量学报, 2015,36(5):550-554.
[7]Ambrosi P. Radiation protection and environmental standards[J]. Metrologia, 2009,46(2):S99-S111.
[8]丁亚东, 魏可新, 宋明哲, 等. 荧光X射线参考辐射的优化[J]. 辐射防护, 2014,34(6):366-369.
[9]闻婷婷, 谭大刚. K荧光辐射强度与管电压关系实验研究[J]. 沈阳师范大学学报(自然科学版), 2007,25(3):323-324. |
|
|
|