Co-planar Array Capacitive Sensor Output Signal Quality Evaluation |
LI Ruihang,ZHANG Yuyan,WEN Yintang,PAN Zhao,LI Hongyang |
Key Lab of Measurement Technology and Instrumentation of Hebei Province, Yanshan University, Qinhuangdao, Hebei 066004, China |
|
|
Abstract The output signal quality of a 12-electrode co-planar capacitive sensor (CCS) for each electrode pair was systematically evaluated by finite element simulations and experiments.The results showed that the output signal quality of the electrode pairs was mainly related to the electrode spacing and the contact edge length.Reducing the electrode spacing and increasing the contact edge length are beneficial for improving the output signal quality.An electrode utilization rate (EUR) method was proposed to evaluate the output signal quality of CCS.The output signal quality of CCS with different number of independent electrodes was evaluated by combining EUR.According to the findings,adding more independent electrodes barely improved the useful output information and generated a significant amount of noise.The EUR value declines as the number of independent electrodes rises,which implies that adding more independent electrodes lowers the output signals quality.The outcomes of the image reconstruction were in agreement with the results of the EUR analysis.The effectiveness of EUR for evaluating the output signal quality of multi-electrode CCS was verified experimentally and in simulation.The methods and conclusions may provide a reference for CCS array electrode design and output signal processing.
|
Received: 21 March 2023
Published: 06 June 2024
|
|
Fund:S&T Program of Hebei;S&T Program of Hebei;S&T Program of Hebei;Natural Science Foundation of Hebei Province;S&T Program of Hebei |
|
|
|
[2] |
VOSS A, HOSSEINI P, POUR-GHAZ M, et al.Three-dimensional electrical capacitance tomography-A tool for characterizing moisture transport properties of cement-based materials [J]. Materials & Design, 2019, 181 (5): 107967.
|
[4] |
YIN X, HUTCHINS D A.Non-destructive evaluation of composite materials using a capacitive imaging technique [J]. Composites Part B: Engineering, 2012, 43 (3): 1282-1292.
|
[6] |
CHENG H, ZHANG Y, HUANG X, et al.Analysis of a concentric coplanar capacitor for epidermal hydration sensing [J]. Sensors & Actuators A Physical, 2013, 203: 149-153.
|
[7] |
KORKUA S K, SAKPHROM S.Low-cost capacitive sensor for detecting palm-wood moisture content in real-time [J]. Heliyon, 2020, 6 (8): e04555.
|
[8] |
LI Z, YIN X K, YAN A, et al.Characterizing Surface Features on Conducting Specimens Through an Insulation Layer Using the Capacitive Imaging Technique [J]. AIP Conference Proceedings, 2016, 1706 (1): 1-4.
|
[10] |
DIAMOND G G, HUTCHINS D A, GAN T H, et al.Single-sided capacitive imaging for NDT [J]. Insight-Non-Destructive Testing and Condition Monitoring, 2006, 48 (12): 724-730.
|
[1] |
FASCHING G E, LOUDIN W J.Capacitive system for three-dimensional imaging of fluidized-bed density [J]. IEEE Transactions on Instrumentation & Measurement, 1994, 43 (1): 56-62.
|
[5] |
JIAO J, LI L, WU B, et al.Novel capacitive proximity sensors for assessing the aging of composite insulators [J]. Sensors and Actuators A: Physical, 2017, 253: 75-84.
|
[11] |
GEBRIAL W, PRANCE R J, HARLAND C J, et al.Noninvasive imaging using an array of electric potential sensors [J]. Review of Scientific Instruments, 2006, 77 (6): 063708.
|
[15] |
ABDOLLAHI-MAMOUDAN F, SAVARD S, IBARRA-CASTANEDO C, et al.Influence of different design parameters on a coplanar capacitive sensor performance [J]. NDT & E International, 2022, 126 (1): 102588.
|
[18] |
ZHANG Y Y, SUN Y R, WEN Y T.An imaging algorithm of planar array capacitance sensor for defect detection [J]. Measurement, 2021, 168 (15): 108466.
|
[21] |
YANG W.Design of electrical capacitance tomography sensors [J]. Measurement Science and Technology, 2010, 21 (4): 042001.
|
[3] |
MOROZOV M, JACKSON W, PIERCE S G.Capacitive imaging of impact damage in composite material [J]. Composites Part B: Engineering, 2017, 113 (15): 65-71.
|
[13] |
GARIMELLA S V, TAYLOR S H.Design of electrode arrays for 3D capacitance tomography in a planar domain [J]. International Journal of Heat & Mass Transfer, 2017, 106: 1251-1260.
|
[23] |
LI R, ZHANG Y, WEN Y, et al.A threshold optimization strategy for reducing the noise of co-planar array capacitive imaging [J]. NDT & E International, 2023, 133 (1): 102738.
|
[9] |
YIN X, HUTCHINS D A, DIAMOND G G, et al.Non-destructive evaluation of concrete using a capacitive imaging technique: Preliminary modelling and experiments [J]. Cement & Concrete Research, 2010, 40 (12): 1734-1743.
|
[19] |
YIN X, HUTCHINS D A, CHEN G, et al.Detecting surface features on conducting specimens through an insulation layer using a capacitive imaging technique [J]. NDT & E International, 2012, 52 (1): 157-166.
|
[12] |
YE Z, BANASIAK R, SOLEIMANI M.Planar array 3D electrical capacitance tomography [J]. Insight-Non-Destructive Testing and Condition Monitoring, 2013, 55 (12): 675-680.
|
[16] |
WEN Y, ZHANG Z, ZHANG Y, et al.Redundancy Analysis of Capacitance Data of a Coplanar Electrode Array for Fast and Stable Imaging Processing [J]. Sensors, 2018, 18 (1): 31.
|
[17] |
ZHANG Y, ZHAO Z, YAO X, et al.Planar array capacitive imaging method based on data optimization [J]. Sensors and Actuators A: Physical, 2022, 347 (1): 113941.
|
[22] |
PENG L, YE J, LU G, et al.Evaluation of Effect of Number of Electrodes in ECT Sensors on Image Quality [J]. IEEE Sensors Journal, 2012, 12 (5): 1554-1565.
|
[14] |
WEI H Y, QIU C H, SOLEIMANI M.Evaluation of planar 3D electrical capacitance tomography: from single-plane to dual-plane configuration [J]. Measurement Science and Technology, 2015, 26 (6): 065401.
|
[20] |
PAN Z, SHAN W A, A P L, et al.An optimization method of planar array capacitance imaging [J]. Sensors and Actuators A: Physical, 2021, 327 (15): 112724.
|
|
|
|