Study on Tube Voltage Value Transfer Technology of Medical Breast X-ray Machine |
REN Guoyue,GUO Siming,QIN Feng,MA Boxuan |
National Institute of Metrology,Beijing 100029, China |
|
|
Abstract With the continuous improvement of the diagnostic X-ray quality evaluation system, tube voltage has become the important index for evaluating the quality of X-ray radiation beams . The measurement method and traceability method of medical breast tube voltage are mainly studied. Two methods are used to measure the tube voltage, namely the voltage divider box voltage division method and the energy spectrum endpoint method.The measurement results are quantitatively transmitted to several semiconductor detectors commonly used in the market. The experimental results show that the maximum difference between the two methods of tube voltage measurement is within 1.0%,the results of measuring the tube voltage using the above two methods have been verified to be reliable. The voltage measurement results of four semiconductor detector tubes commonly used in the domestic market are calibrated, and the deviation of the relative inherent error is within 2.0%, which can be used for on-site quality control of medical breast machines.
|
Received: 19 October 2023
Published: 18 December 2024
|
|
|
|
|
[1] |
杨勇, 张圆月, 刘志宏, 等.医用诊断X射线剂量探测技术概述[J].中国测试技术, 2005, 31(4): 94-96.
|
[7] |
张帅, 吴金杰, 吴冲, 等. HPGe探测器能量刻度及探测效率模拟[J]. 核电子学与探测技术, 2016, 36(5): 517-520
|
[2] |
杨建, 侯新生, 郑永明, 等.基于非介入式X射线管电压测量的研究[J].中国测试技术, 2008, 34(5): 110-112。
|
|
WU R. Research on Component Performance of Resistance Dividing Bias Amplifier[J]. Electronics World, 2019(14): 51-52.
|
|
LIU X, GU J, WANG J J, et al. Design of 10kV DC High Voltage Standard Voltage Divider[J]. Measurement Technique,2002(1): 17-19.
|
[8] |
TOMAS U. Method and apparatus for measuring X-ray radiation:US5761270[P]. 1998-06-02.
|
|
NIST, USA and the BIPM in Mammography X-Rays[J].Metrologia, 2011,48:06014.
|
[3] |
吴瑞. 电阻分压式偏置放大电路元器件性能研究[J]. 电子世界, 2019(14): 51-52
|
[4] |
刘星, 顾杰, 王建军, 等. 10kV直流高压标准分压器的设计[J]. 计量技术, 2002(1): 17-19。
|
[10] |
REBECCA L S, KIMBERLY D M, AHMEDIN J D. Cancer Statistics, 2019[J]. CA: A Cancer Journal for Clinicians, 2019, 69(1): 7-34.
|
[12] |
KESSLER C, BURNS D T, BRIEN M.O. Key Comparison BIPM. RI(I)-K7 of the Air-Kerma Standards of the
|
|
YANG Y, ZHANG Y Y, LIU Z H, et al. Overview of Medical Diagnostic X-ray Dose Detection Technology[J]. China Measurement & Test, 2005, 31(4): 94-96.
|
|
YANG J, HOU X S, ZHENG Y M, et al. Research on Voltage Measurement Based on Non-intrusive X-ray Tube[J]. China Measurement & Test, 2008, 34(5): 110-112.
|
[5] |
MATSUMOTO C, TAKAHASHI T. Performance of a New Schottky CdTe Detector for Hard X-ray Spectroscopy[J]. IEEE Transactions on Nuclear Science, 1998, 45(3): 428-432.
|
[6] |
ROSELI K, HERDADE S B, TERINI R A, et al. X-ray spectroscopy in mammography with asilicon PIN photodiode with application to the measurement of tube voltage[J]. Medical Physics, 2003(11): 2996-3003.
|
|
ZHANG S, WU J J, WU C, et al. Simulation of Energy Scale and Detection Efficiency of HPGe Detector[J]. NUCLEAR ELECTRONICS & DETECTION TECHNOLOGY, 2016, 36(5): 517-520
|
[9] |
McGuire S. World Cancer Report 2014.Geneva, Switzerland:World Health Organization, International Agency for Research on Cancer, WHO Press, 2015 [J]. Advances in Nutrition,2016,7(2): 418-419..
|
[13] |
KRAMER H M, SELBACH H J. Extension of the range of definition of the practical peak voltage up to 300kV[J].British Journal of Radiology, 2008, 81(969):693-698.
|
[14] |
Medical electrical equipment-Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology:IEC 61676[S].2023
|
[16] |
XU M, SUN J, LEE F C. Voltage divider and its application in the two-stage power architecturet [C]//IEEE. Twenty-First Annual IEEE Applied Power Electronics Conference and Exposition.Dallas, TX, USA, 2006: 499-505.
|
[11] |
KESSLER C, BURNS D T, TANKAKA T, et al. Key Comparison BIPM. RI(I)-K7 of the Air-Kerma Standards of the NMIJ, Japan and the BIPM in Mammography X-Rays[J].Metrologia, 2010,47:06023.
|
[15] |
HEINTZ P, CHAMBERS G, SANDOVAL D . SU-E-I-86: Evaluation of the New RaySafe Unfors X2 Dosimetry System[J].Medical Physics, 2014,41(6):150.
|
[17] |
CREAGH D C, HUBBELL J H. Problems associated with the measurement of X-ray attenuation coefficients. I. Silicon. Report of the International Union of Crystallography X-ray Attenuation Project[J].Acta Crystallographica Section A: Foundations and Advances,1987,43(1):102-112.
|
|
|
|