Study on the Measurement of Low Energy X-ray Air Kerma Based on Proportional Counter |
QIN Feng1,2,WU Jin-jie2,LI Meng-yu2,3,LU Ping-zhou2,3,QU Bing-bing2,3,JIAO Zhi-wei1,TANG Xian-qiang2,3,ZHU Wei-feng2,3,ZHAO Rui2 |
1. China Jiliang University, Hangzhou, Zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China
3. Chengdu University of Technology, Chengdu, Sichuan 610059, China |
|
|
Abstract In order to accurately measure low-energy and low-dose-rate X-ray, based on 10~60kV X-ray machine, seven reference radiation quality were established by adding additional filtration on the basis of low air kerma series radiation quality. The energy spectrum of radiation quality by Monte Carlo simulation program were obtained, and half-value layer and average energy were calculated. The proportional counter was used to complete the measurement of the energy spectrum. The results showed that the relative deviation between the simulated energy spectra and measured spectra does not exceed 6%. By used PTW34047 ionization chamber, the measurement of air kerma rate was completed by successive substitution method. The proportional counter was calibrated and the energy spectrum-dose conversion function was obtained. Finally, the energy spectrum-dose conversion function was verified, and the deviation from the results obtained by the ionization chamber method is not more than 3.8%. Verified by AT1123 survey meter, the relative deviation is within 9%.
|
Received: 25 December 2022
Published: 22 August 2023
|
|
|
|
|
[2] |
姚叶豹, 马得勋, 殷爱民,等. 低剂量电离辐射对人体健康影响的研究现状 [J]. 辐射防护通讯, 2020, 40(3): 1-8.
|
|
Zhao R, Wu J J, Fan S, et al. Establishment of environmental level X-ray radiation qualities and research of the calibration methods of monitoring instruments [J]. Acta Metrologica Sinica, 2018, 39(S1): 131-136.
|
[3] |
彭峰莉. 辐射环境监测在辐射环境影响评价中的应用分析 [J]. 环境与发展, 2019, 31(11): 130-131.
|
|
Environment and Development, 2019, 31(11): 130-131
|
[5] |
Beddingfield D H, Beyerle A, Russo P A, et al.
|
|
High-Preesure xenon ion chambers for gamma-ray spectroscopy in nuclear safeguards[J]. Nuclear Instruments and Methods in Physics Research A, 2003, 505(1/2): 474-477.
|
[7] |
张勇, 杨勇. 一种改善闭气式正比计数器换气周期的方法 [J]. 核电子学与探测技术, 2011, 31(11): 1292-1294.
|
[11] |
张江云, 刘艳芳, 刘鹏, 等. 正比计数管探测低能X射线时探测效率的MC计算[J]. 核电子学与探测技术, 2016, 36 (3): 312-315.
|
[12] |
郭思明, 刘莹, 吴金杰, 等. 乳腺 X 射线基准装置半值层的测量 [J]. 计量学报, 2018, 39 (3): 410-413.
|
[23] |
GBT 12162.3-2004 用于校准剂量仪和剂量率仪及确定其能量响应的 X和γ参考辐射第3部分: 场所剂量仪和个人剂量计的校准及其能量响应和角响应的测定[S]. 2004.
|
[4] |
Akimov D Y, Burenkov A A, Churakov D L,et al. Development of High Pressure Xe Scintillation Proportional Counter for Experiments in “Low-Background” Physics[J]. Physics,1997,doi:10.48550/arXiv.hep-ex/9703011.
|
|
Zhang Y, Yang Y. A Method for Optimising the Gas Replacing Period of Sealed Proportional Counter [J]. Nuclear Electronics & Detection Technology, 2011, 31(11): 1292-1294.
|
|
Li M Y, Zhao R, Qu B B,et al. Establishment of low-dose rate X-ray radiation quality and calibration of the instruments [J]. Nuclear Techniques, 2022, 45(6): 61-67.
|
|
Song F, Zhao R, Ding W C,et al. Research on calibration method of X-ray environmental radiation monitoring instrument [J]. Acta Metrologica Sinica, 2021, 42(9): 1237-1243.
|
[10] |
鲁平周, 吴金杰, 李梦宇,等. 低能X射线水吸收剂量测量的散射辐射研究 [J]. 核技术, 2022, 45(4): 33-40.
|
|
Guo S M, Liu Y, Wu J J, et al. Half Value Layer Measurem ent of Mammography X-ray Standard [J]. Acta Metrologica Sinica, 2018, 39(3): 410-413.
|
|
Wang P, Xing L T, Jiang W. Measurement and Research for the Filtered X-ray Radiation Quality of Low Air Kerma Rate Series [J]. Acta Metrologica Sinica, 2021, 42(4): 521-525.
|
|
Wang J W, Wei P, Ma B X, et al. Main Factors Affecting Measurement Results of the Half-value Layer of Diagnostic X-ray Testing Equipment [J]. Metrology Science and Technology, 2021, 65(8): 3-6+28.
|
[15] |
张德亮, 吴金杰, 杜海燕,等. 低空气比释动能率X射线参考辐射质的建立 [J]. 核电子学与探测技术, 2018, 38(2): 217-222.
|
[22] |
黄平, 万清, 董文龙, 等. 基于能谱-剂量转换函数的 NaI 辐射探测器能量响应修正方法研究 [J]. 中国测试, 2022, 48(2): 124-128.
|
|
Yao Y B, Ma D X, Yin A M,et al. Research Progress on Effects of Low Dose Ionizing Radiation on Human Health [J]. Radiation Proctection Billetin, 2020, 40(3): 1-8.
|
|
Peng F L. Application of radiation environmental monitoring in radiation environmental impact assessment [J].
|
[8] |
李梦宇, 赵瑞, 屈冰冰,等. 低剂量率X射线辐射质的建立与仪器校准 [J]. 核技术, 2022, 45(6): 61-67.
|
[9] |
宋飞, 赵瑞, 丁卫撑,等. X射线环境辐射监测仪器校准方法的研究 [J]. 计量学报, 2021, 42(9): 1237-1243.
|
[13] |
王鹏, 邢立腾, 蒋伟. 低空气比释动能率系列过滤 X射线辐射质的测量与研究 [J]. 计量学报, 2021, 42(4): 521-525.
|
[14] |
王家伟, 魏鹏, 马伯轩, 等. 诊断 X 射线检测设备半值层测量结果的影响因素 [J]. 计量科学与技术, 2021, 65(8): 3-6+28.
|
[21] |
曹蕾, 李德红, 杨扬, 等. 能谱-剂量转换法用于环境辐射剂量测量 [J]. 同位素, 2022, 35(4): 297-303.
|
[1] |
赵瑞, 吴金杰, 樊松, 等. 环境水平X射线辐射质的建立及监测仪表校准方法研究 [J]. 计量学报, 2018, 39(S1): 131-136
|
[6] |
Jang D Y, Park B H, Kang S Y, et al. A Cylindrical High-Pressure Xenon Chamber with Shielding Mesh[C]//IEEE. Nuclear Science Symposium Conference Record (NSS/MIC). Orlando, FL, 2009.
|
|
Zhang J Y, Liu Y F, Liu P, et al. Efficiency Monte Carlo Calculation of Proportional Counter Tube for Low-energy X-Ray Detecting [J]. Nuclear Electronics & Detection Technology, 2016, 36 (3): 312-315.
|
[16] |
Bandalo V, Greiter M B, Brnner J, et al. ISO 4037: 2019 validation of radiation qualities by means of half-value layer and Hp(10) dosimetry [J]. Radiation Protection Dosimetry, 2019, 187(4): 438-450.
|
[17] |
段小娟, 吴金杰, 杨元第. EGSnrc程序模拟低能X射线光管参数对射线能谱的影响 [J]. 计量学报, 2011, 32(S1): 51-53.
|
[18] |
Dombrowski H, Neumaier S. Traceability of the PTB low-dose rate photon calibration facility [J]. Radiation Protection Dosimetry, 2010, 140(3): 223-233.
|
[19] |
Neumaier S, Arnold D, Bohm J, et al. The PTB underground laboratory for dosimetry and spectrometry [J]. Applied Radiation & Isotopes, 2000, 53(1): 173-178.
|
|
Lu P Z, Wu J J, Li M Y,et al. Study on scatter radiation of low energy X-ray absorbed dose to water measurement [J]. Nuclear Techniques, 2022, 45(4): 33-40.
|
|
Zhang D L, Wu J J, Du H Y,et al. Establishment of X-ray Reference Radiation of Low Air-kerma Rate [J]. Nuclear Electronics & Detection Technology, 2018, 38 (2): 217-222.
|
|
Huang P, Wan Q, Dong W L, et al. Research on energy response correction method for NaI radiation detector based on spectrum-dose conversion function [J]. China Measurement & Test, 2022, 48(2): 124-128.
|
|
Duan X J, Wu J J, Yang Y D. The affect of X-rays tube parameters simulated by EGSnrc on low energy spectrum [J]. Acta Metrologica Sinica, 2011, 32(S1): 51-53
|
[20] |
Neumaier S, Wojcik M, Dombrowski H, et al. Improvements of a low-level gamma-ray spectrometry system at the underground laboratory “UDO” [J]. Applied Radiation & Isotopes Including Data Instrumentation & Methods for Use in Agriculture Industry & Medicine, 2009, 67(5): 726-730.
|
|
Cao L, Li D H, Yang Y et al. Enery Spectrum to Dose Conversion Method Used in Environment Radiation Dose Measurement [J]. Journal of Isotopes, 2022, 35(4): 297-303.
|
|
|
|