|
|
Calibration of 1mT~2.5T Magnetometers |
XU Yu1, CAO Yan2, CHEN Gui-ying1, ZHOU Xin-hua3 |
1.Hunan Institute of Metrology and Test, Changsha, Hunan 410014, China
2.Changsha Aviation Vocational and Technical College,Changsha, Hunan 410124, China
3.Changsha Tunkia Measurement and Control Technology Co. Ltd., Changsha, Hunan 410100, China |
|
|
Abstract In order to meet the calibration requirements of magnetometers ranging from 1mT to 2.5T, a magnetometer calibration device is designed, The constant magnetic field of the device is generated by magnetic field coil and electromagnet. The position adjustment device of the probe is controlled by computer, which can ensure that the Hall plate of the probe is perpendicular to the direction of the magnetic field. The whole measurement process is controlled by computer, which is convenient and fast. At the same time, the calibration method of the main technical parameters of magnetometer is expounded. The calibration method in the magnetic field of electromagnet is described in detail, which solves the difficult problem of magnetometer calibration. Finally, through calibrating digital magnetometer, the expanded uncertainty at the measurement point of magnetic field intensity 1T is 0.92mT.
|
Received: 12 September 2019
|
|
|
|
|
1 TumanskiS. 磁性测量手册[M]. 北京: 机械工业出版社, 2014, 112-119.
2 王锐, 袁静. 一种特斯拉计自动校准系统设计[J]. 计量与测试技术, 2012, 39(4): 45-46. WangR, YuanJ. The Design of Tesla-Meter Automatic Calibration System[J]. Metrology & Measurement Technique, 2012, 39(4):45-46.
3 杜胜雪, 孔令富, 李英伟. 电磁流量计矩形与鞍状线圈感应磁场的数值仿真[J]. 计量学报. 2016, 37(1): 38-42. DuS X, KongL F, LiY W. Numerical simulation of the induced Magnetic Field of Rectangular and Saddle Coils in Electromagnetic Flowmeter[J]. Acta Metrologica Sinica. 2016, 37(1): 38-42.
4 国家质量技术监督局. JJG 242—1995特斯拉计检定规程[S]. 1995.
5 张勤, 曹瑞基. 电磁计量[M]. 北京: 中国计量出版社, 2007: 415-425.
6 李玉石. 磁参量计量标准装置的设计和建立 [J]. 计量与测试技术, 2013, 40(8): 12-13. LiY S. Design and Development of Magnetic Standard Facility[J]. Metrology & Measurement Technique, 2013, 40(8): 12-13.
7 李鑫, 张伟. 磁强计校准装置的建立[J]. 电测与仪表, 2012,(10): 43-45. LiX, ZhangW. The Establishment of Magnetometer Calibration Device[J]. Electrical Measurement & Instrumentation. 2012,(10): 43-45.
8 华长春, 马立强. 二等磁参量计量标准的建立[J]. 现代测量与实验室管理. 2011, (3): 6-7. HuaC C, MaL Q. Development of Secondary Magnetic Standard Facility[J]. Advanced Measurement And Laboratory Mangement. 2011, (3): 6-7.
9 倪育才. 实用测量不确定度评定[M]. 北京: 中国计量出版社, 2008.
10 丁莹, 董全林, 赵伟霞, 等. 两类鞍形线圈产生磁场均匀性效能的研究[J]. 计量学报. 2016, 37(6): 638-643. DingY, DongQ L, ZhaoW X, et al. Research on Magnetic uniformity performance of two kinds of Saddle Coils for Magnetic Shielding[J]. Acta Metrologica Sinica. 2016, 37(6): 638-643.
11 何冰心, 王瑞东, 孟凡茂. 霍尔效应法测量亥姆霍兹线圈空间磁场均匀区[J]. 沈阳师范大学学报(自然科学版), 2005, 23(4): 358-360. HeB X, WangR G, MengF M. Measure the uniformity of Helmholtz Coil space Magnetic field with Hall effect method[J]. Journal of Shenyang Normal University(Natural Science), 2005, 23(4): 358-360.
12 邹志纯. 亥姆霍兹线圈空间的磁场分布[J]. 西安邮电学院学报, 2004, 9(3): 89-91. ZouZ C. The distribution of the Magnetic Field intensity of the Helmholtz Coil[J]. Journal of Xi'an University of Post and Telecommunication, 2004, 9(3): 89-91. |
|
|
|