Numerical Simulation of Gas-liquid Two-phase Flow in Coriolis Flowmeter |
MA Ya-peng1,SHI Chu1,LUO Yu1,SUGI Tokio2,PANG Qiang3,AKIZAKI Tsubasa3 |
1.School of Naval Architecture, Ocean and Civil Engineering, Shanghai Jiaotong University, Shanghai 200240, China
2.Tokyo Keiso Co. Ltd., Tokyo 144-0051, Japan
3.Tokyo Keiso Co. Ltd., Shanghai 201612, China |
|
|
Abstract In the measurement of gas-liquid two-phase fluid, the Coriolismass flowmeter (CMF) measurement stability of the flow meter gradually deteriorated with the gas content increases. A method based on the fluid-solid coupling finite element technique has been developed for the modal analysis and fluid-solid coupling vibration analysis of the gas-fluid two-phase flow in Coriolis flowmeter. As an example, this method is used to simulate the effect of bubble content and distribution on measurement stability. The variation law of measurement stability has been preliminary mastered, that the measurement is relatively stable with a bubble content of 1%, and over 5% the stability changes rapidly.
|
Received: 28 June 2017
Published: 05 September 2018
|
|
|
|
|
[1]Anklin M, Drahm W, Rieder A.Coriolis mass flowmeters: Overview of the current state of the art and latest research[J]. Flow Measurement and Instrumentation, 2006, 17(6):317-323.
[2]Gregory D, West M, Paton R, et al. Two-phase flow metering using a large Coriolis mass flow meter applied to ship fuel bunkering[J]. Measurement and Control, 2008, 41(7): 208-212.
[3]Seeger M. Coriolis flow measurement in two phase flow[J]. IEEE Computing & Control Engineering,2005, 16 (3): 10-16.
[4]Reizner J. Coriolis-the almost perfect flow meter[J].IEEE Computing & Control Engineering, 2003, 14 (4):28-33.
[5]Hemp J, Yeung H. Coriolis meters in two phase conditions[J]. IEEE Computing & Control Engineering, 2003, 14(4): 36.
[6]Liu R P, Fuent M J, Henry M P, et al.A neural network to correct mass flow errors caused by two-phase flow in a digital Coriolis mass flowmeter[J]. Flow Measurement and Instrumentation, 2001, 12(1): 53-63.
[7]Henry M P, Tombs M, Duta M, et al. Two-phase flow metering of heavy oil using a Coriolis mass flow meter: A case study[J]. Flow Measurement and Instrumentation, 2006, 17(6) : 399-413.
[8]Henry M P, De La Fuente M J. Correcting for two-phase flow in a digital flowmeter:US 9,021,892[P]. 2015-06-02.
[9]Henry M P, De La Fuente M J. Correcting for Two-Phase Flow in A Digital Flowmeter:US 6, 918, 42[P].2006-01-03.
[10]Liu R P, Fuent M J, Henry M P, et al. A neural network to correct mass flow errors caused by two-phase flow in a digital Coriolis mass flowmeter[J]. Flow Measure & Instrumentation, 2001,12(1): 53-63.
[11]董帅, 徐科军, 侯其立, 等. 微弯型科氏质量流量计测量气-液两相流研究[J]. 仪器仪表学报, 2015, 36(9): 1972-1977.
[12]Sultan G, Hemp J. Modelling of Coriolis Mass Flowmeter[J]. Journal of Sound and Vibration,1989, 32(3):473-489.
[13]Bobovonk G, Mole N, Kutin J, et al. Coupled finite-volume/finite-element modeling of the straight-tube Coriolis flow meter[J]. Journal of Fluids and Structures, 2005, 20: 785-800.
[14]Mole N, Bobovonk G, Kutin J, et al.An improved three-dimensional coupled fluid-structure model for Coriolis flowmeters[J]. Journal of Fluids and Structures, 2008, 24: 559-575.
[15]刘颖,罗宇,杉时夫,等.基于流固耦合计算的科里奥利质量流量计计算机辅助设计[J]. 船舶工程, 2013, S1: 140-143.
[16]朱小倩.科氏流量计两相流测量含气率影响规律及修正方法的研究[D].青岛: 中国石油大学(华东), 2011: 12-13.
[17]金艳,王利军,胡亮,等. 科氏流量计测量管固有频率在线预测方法[J]. 计量学报, 2017, 38(3): 328-332. |
|
|
|