The Realization of 380~780 nm Total Spectral Radiant Flux and the Measurement Uncertainty Evaluation |
ZHAO Wei-qiang,LIU Hui,LIU Jian |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract The basic principle of the realize 380~780 nm total spectral radiant flux (TSRF) scale is described, based on measuring total luminous flux and measuring relative spectral radiant intensity distribution by a goniospectroradiometer. The TSRF quantity traces to the primary standard of total luminous flux, and the primary standard of spectral radiant intensity, the high temperature black body. The measurement uncertainty components of TSRF scale include the total luminous flux, spectral radiant intensity standard lamp, the array spectroradiometer, the goniospectroradiometer (mechanics) and the repeatability. The expanded uncertainty is evaluated to be U=3.3%~1.6%, k=2。
|
Received: 04 January 2017
Published: 29 December 2017
|
|
|
|
|
[1]刘慧,赵伟强,刘建,等.LED 总光谱辐射通量测量技术的研究[J].计量学报,2014,35(6):574-577.
[2]中国计量科学研究院自动分布光度计课题组.复合式大型自动分布光度计[J].计量学报,1983,4(4):250-255.
[3]代彩红,吴志峰,欧阳慧泉,等.第四代光谱辐射度和色温度国家基准装置的研制[J].计量学报,2013,34(3):201-206.
[4]Zong Y, Ohno Y. Realization of total spectral radiant flux scale and calibration service at NIST[C]//CIE. 26th Session of the CIE. Beijing, 2007, D2: 179-182.
[5]Shaw M, Goodman T. Array-based goniospectroradiometer for measurement of spectral radiant intensity and spectral total flux of light sources [J]. Applied Optics, 2008, 47(14):2637-2647.
[6]Godo K, Niwa K, Kinoshita K, et al. Realization of total spectral radiant flux scale at NMIJ with a goniophotometer/spectroradiometer[J].Metrologia, 2016,53(2):853-859.
[7]代彩红,吴志峰,王彦飞,等.NIM和VNIIOFI光谱辐射照度双边比对[J].计量学报,2016,37(4): 347-351.
[8]Zong Y, Brown S W, Johnson B C, et al. Simple spectral stray light correction method for array spectroradiometers[J]. Applied Optics, 2006,45(6):1111-1119.
[9]赵伟强,刘慧,刘建,等.采样间隔对分布光度计测量LED总光通量的影响[J].计量学报,2015,36(1):22-25. |
|
|
|