The Development of Axial Standard Device for Calibration of Ophthalmometers |
ZHANG Ji-yan,LI Jiao,GAO Ming-liang,DING Xiang |
Division of Medical and Biological Measurement, National Institute of Metrology, Beijing 100029, China |
|
|
Abstract In order to resolve the axis calibration of ophthalmometers, aixal standard device is developed which consists of axial model eye, axial sleeve and measurement support. Axial model eye is a square cylinder whose front surface is toroidal and back surface is scrub plane, which is located in the square through hole of axial sleeve, and axial sleeve is located on the trapezoid groove of measurement support. By assembly and mechanical positioning,the measurement of four axes of 0°, 45°, 90° and 135° are achieved and axis uncertainty is about 0.5° with coverage factor of 2, which meets the requirement of axis calibration for different kinds of ophthalmometers. Moreover, the developed axial standard device not only expands measurement capabilities of the existing working standard of ophthalmometers, but also provides technical support for revision of related verification regulations.
|
Received: 11 June 2015
Published: 19 April 2017
|
|
Corresponding Authors:
ZHANG JiJiyan
E-mail: zhangjy@nim.ac.cn
|
|
|
|
[1]张吉焱, 王莉茹, 陈燕. 角膜曲率计计量标准器的研制[J]. 计量学报, 2006, 27(1): 28-31.
[2]张吉焱,洪宝玉,高明亮. 样板检验法测量角膜曲率计用计量标准器的不确定度评定[J]. 计量技术,2013,(9):47-50.
[3]国家质量监督检验检疫总局. JJG 1011—2006 角膜曲率计[S]. 2006.
[4]国家质量监督检验检疫总局. JJG 1088—2013 角膜曲率计用计量标准器[S]. 2013.
[5]ISO. ISO 10343—2014 Ophthalmic instruments—Ophthalmometers[S]. 2014.
[6]吴燮灿. 实用眼镜光学[M]. 北京:北京科学技术出版社,2007: 21-103.
[7]ISO. ISO 13666: 2012 Ophthalmic optics—Spectacle lenses—Vocabulary[S]. 2012.
[8]GB/T 26397—2011 眼科光学 术语[S]. 2011.
[9]李士贤,李林. 光学设计手册:修订版[M]. 北京:北京理工大学出版社,1996.
[10]张以谟. 应用光学[M]. 北京:机械工业出版社,1982. |
|
|
|