Solution of Building the Standard Laboratory of Radon |
ZHANG Yang,XU Ming,SUN Gao-feng,YIN Guo-hui,HU Mao-zhong, ZOU Rong-hu,WANG Yin-dong,CHENG Zhi-wei,FENG Tian-cheng |
Northwest Institute of Nuclear Technology, Xi’an, Shaanxi 710024, China |
|
|
Abstract The standard laboratory of radon plays an important role in the tracing and transmitting of value of radon concentration. But there are many problems in the process of building the standard laboratory of radon, which are gas leaking of solid radon source, inconformity between theoretical value and experimental value of radon concentration in laboratory, and crystallization of liquid radon source. Measures were taken to solve these problems, such as changing the gas washing canal of radon, optimizing the model of gas washing, multiple sealing and filtering, building empirical curve, and so on. It indicated that the relative deviation between theoretical value and experimental value of radon concentration in laboratory was less than ±3.8%, and the long time stability of liquid radon source less than 0.7%, which meet the requirment of measurment and methods established metrology.
|
Received: 20 July 2015
Published: 28 February 2017
|
|
|
|
|
[1]潘自强.电离辐射环境监测与评价[M].北京: 原子能出版社, 2007.
[2]Charles M. UNSCEAR report 2000: Sources and effects of ionizing radiation[J]. Protection Offical Journal of the Society for Radiological Procection, 2001, 21(1):83-85 .
[3]张洋,尹国辉,吴睿,等. 就地γ谱仪测量某坑道内222Rn-220Rn子体剂量比[J].核电子学与探测技术, 2012, 32 (4) : 433-437.
[4]李宗扬.计量技术基础[M].北京: 原子能出版社, 2002.
[5]荣超凡.电离辐射计量[M].北京: 原子能出版社, 2002.
[6]卢希庭.原子核物理[M].北京: 原子能出版社, 2000.
[7]JJG 825—1993, 测氡仪[S]. |
|
|
|