Acta Metrologica Sinica  2017, Vol. 38 Issue (1): 117-122    DOI: 10.3969/j.issn.1000-1158.2017.01.25
Current Issue | Archive | Adv Search |
The Establishment of the On-site Calibration Device for 2πα and 2πβ Particle Emission Rate Measurement
LI Ze-xi1,2,HE Hai-jiang2,3,LIANG Jun-cheng2,SHAN Jian1,GAN Ya-zhou4,YANG Zhi-jie2,3,LIU Hao-ran2,ZHANG Ming2
1. School of Nuclear Science and Technology of University of South China, Hengyang, Hunan 421001, China
2. National Institute of Metrology, Beijing 100029, China
3. China University of Geosciences, Beijing 100083, China
4. CGN(Shenzhen) Radiation Monitoring Technology Co Ltd, Shenzhen, Guangdong 518124, China
Download: PDF (936 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  Based on a large-area, windowless, gas-flow proportional counter, an alpha/beta 2π emission rate on-site calibration device is established. Performance tests are taken and the results show that the plateau of alpha lasts from 700 V to 1 500 V, the length is up to 800V and the slope is 0.25% per 100V, the plateau of beta lasts from 1650V to 2050V, the length is up to 400V and the slope is 0.44% per 100V. This on-site calibration device is used to measure the standard surface sources which are calibrated by the the absolute alpha/beta 2π emission rate measuring system, the results are well agreed with the standard values within the range of uncertainties.On-site measurement of another group of standard surface sources are taken, values of corrected results are agreed with the values calibrated in previous cycles within the range of uncertainties, which verifies the feasibility of this calibration device to deliver the on-site standard values.
Key wordsmetrology      particle emission rate      plane source      standard value transfer      nuclides contamination     
Received: 01 June 2016      Published: 28 December 2016
PACS:  TB98  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
LI Ze-xi
HE Hai-jiang
LIANG Jun-cheng
SHAN Jian
GAN Ya-zhou
YANG Zhi-jie
LIU Hao-ran
ZHANG Ming
Cite this article:   
LI Ze-xi,HE Hai-jiang,LIANG Jun-cheng, et al. The Establishment of the On-site Calibration Device for 2πα and 2πβ Particle Emission Rate Measurement[J]. Acta Metrologica Sinica, 2017, 38(1): 117-122.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2017.01.25     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2017/V38/I1/117
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech