Acta Metrologica Sinica  2016, Vol. 37 Issue (2): 214-218    DOI: 10.3969/j.issn.1000-1158.2016.02.23
Current Issue | Archive | Adv Search |
Monte Carlo Simulation and Experimental Validation of Correction Factor for Cylindrical Cavity Ionization Chamber
WU Meng-meng1,LI De-hong2,WANG Pei-wei2
1. China Institute of Atomic Energy, Beijing 102413, China
2. National Institute of Metrology, Beijing 100029, China
Download: PDF (1154 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  To discuss how to use nominal 10 cm3 cylindrical cavity ionization chamber to give absolute measurement to the Cs-137 air kerma. The wall correction factor kwall and the axial non-uniformity correction factor kan is determined by both the experimental method and Monte Carlo (MC) calculation. The results showed, the wall correction factor kwall determined by geometry thickness extrapolation method is less than 0.091% compared with that by the MC calculation, the axial non-uniformity correction factor kan determined by the experimental method is less than 0.63% compared with that by the MC calculation. The error between the two values is acceptable. The repeated experiment will inevitably introduce other error factors, and MC calculation method is simple and high repeatability. Eventually we selected MC calculation results for kwall, kan value.
Key wordsmetrology      air kerma      cavity ionization chamber      wall correction factor      axial non-uniform correction factor      Monte Carlo     
Received: 12 June 2015      Published: 27 January 2016
PACS:  TB98  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
WU Meng-meng
LI De-hong
WANG Pei-wei
Cite this article:   
WU Meng-meng,LI De-hong,WANG Pei-wei. Monte Carlo Simulation and Experimental Validation of Correction Factor for Cylindrical Cavity Ionization Chamber[J]. Acta Metrologica Sinica, 2016, 37(2): 214-218.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2016.02.23     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2016/V37/I2/214
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech