Study on Approximating the ITS-90 Temperature Scale with Industrial Platinum Resistance Thermometers |
JIN Zhi-jun,ZHANG Jin-tao,ZHANG Zhe |
National Institute of Metrology,Beijing 100013,China |
|
|
Abstract For industrial-type platinum resistance thermometer (IPRTs), the most-used interpolation model (resistance versus temperature ) is based on the Callendar- Van Dusen (CVD) equation,and standard platinum resistance thermometer (SPRT) with interpolation equation based on the ITS-1990 .There are difference between CVD equation and ITS-1990.The experiment results show, it is feasible to use ITS-1990 interpolation methords for IPRTs.There are more uniformity and accuracy with ITS-1990 than that with CVD equation.
|
|
|
|
|
|