Radiance Temperature Characteristics on Wavelength of Tungsten Strip Lamps and Relationship with Lamp Structures |
YUAN Zun-dong |
National Institute of Metrology, Beijing 100013, China |
|
|
Abstract The wavelength correction for radiance temperature is often occurred in the realization, dissemination, and international comparison of the International Temperature Scale of 1990. The relation of the structures and the dependences of radiance temperature on wavelength of tungsten strip lamps is discussed. The derivative of the radiance temperature with wavelength for the T type tungsten strip lamp in 400~1 100 nm is calculated, and tests are carried out for a set of large wavelength interval corrections with the experimental data. The result is helpful to improve the radiance temperature correction and extend the spectral range of the applications of tungsten strip lamps.
|
|
|
|
|
|
[1]Kostkowski H J, Lee R D. Theory and Methods of Optical Pyrometry [A]. Temperature: Its Measurement and Control in Science and Industry[C], New York: 1962.
[2]赵琪, 崔志尚.有效波长理论与亮温跟踪[J].计量学报,1983,4(1):41-51.
[3]原遵东,段宇宁,王铁军,等.复现银凝固点以上国际温标(ITS-90)[J].现代计量技术,2001,9(2):13-21.
[4]原遵东.光电高温计校准中的有效波长修正[J].计量技术, 2008,(5):42-45.
[5]CCT-K5: Realizations of the ITS-90 between 962 ℃ and 1 700 ℃ [EB]. http://kcdb.bipm.org/AppendixB/KCDB_ApB_search_result.asp?search=1&met_idy=9&bra_idy=30&cmt_idy=1&ett_idy_org=0&epo_idy=0&cou_cod=0,2007-12-26.
[6]赵琪,李大今,郎兆林,等.钨带灯亮温-波长相关性实验研究[J].计量学报,1984,5(4):288-294.
[7]Latyev L N, Chekovsky V Y, Shestakov E N. Monochro-matic emissivity of tungsten in the temperature range 1 200~2 600 K and in the wavelength range 0.4~4 μm [J]. High Temp High Press, 1970, 2: 175-181. |
|
|
|