Experimental Study of Flow Measurement Characteristics of Venturi Tube under Pulsating Flow Conditions

ZHANG Xuerong, LUO Haotian, LIU Li, GU Hanyang

Acta Metrologica Sinica ›› 2026, Vol. 47 ›› Issue (4) : 523-530.

PDF(3718 KB)
PDF(3718 KB)
Acta Metrologica Sinica ›› 2026, Vol. 47 ›› Issue (4) : 523-530. DOI: 10.3969/j.issn.1000-1158.2026.04.07

Experimental Study of Flow Measurement Characteristics of Venturi Tube under Pulsating Flow Conditions

Author information +
History +

Abstract

In the operation of many industrial equipment such as rotating equipment, reciprocating equipment and regulating valves, the generation of pulsating flow is often caused, and the unstable flow of fluid has a great influence on the accuracy of flow measurement. By building an experimental circuit for pulsating flow measurement, using Venturi and differential pressure sensor to measure flow and differential pressure under different working conditions, the measurement characteristics of Venturi under pulsating flow are studied, the effects of pulsating flow, lead pipe length and pulsation frequency on the accuracy of differential pressure measurement are explored, and the feasibility of high-frequency pressure sensor to measure the pulsating differential pressure is comparatively analyzed. The results show that the increase of pulsating flow will cause the amplitude of the pulsating differential pressure to increase, the length of the pilot tube and the pulsation frequency have a significant effect on the measurement results of the pulsating differential pressure, and the shorter pilot tube and lower pulsation frequency can help to improve the accuracy of the measurement of the pulsating differential pressure, the high-frequency pressure sensor has the reliability of the measurement under the conditions of pulsating flow, and it can effectively eliminate the error caused by the pilot tube. The results of the study provide a reference for the optimization of fluid measurements under pulsating flow conditions.

Key words

flow measurement / pulsating flow / Venturi tube / pulsating differential pressure / differential pressure sensor / length of pressure tap tube / pulsation frequency

Cite this article

Download Citations
ZHANG Xuerong , LUO Haotian , LIU Li , et al. Experimental Study of Flow Measurement Characteristics of Venturi Tube under Pulsating Flow Conditions[J]. Acta Metrologica Sinica. 2026, 47(4): 523-530 https://doi.org/10.3969/j.issn.1000-1158.2026.04.07

References

[1]
ALESBE I ABDEL-MAKSOUD M ALJABAIR S. Analysis of unsteady flow over offshore wind turbines in combination with different types of foundations[J]. J Marine Sci Appl201716: 199-207.
[2]
张郅豪, 赵佳锡, 章荣平, 等.高压临界流文丘里喷嘴气体流量标准装置温度场测量与分析[J].计量学报202546(3):368-375.
ZHANG Z H ZHAO J X ZHANG R P, et al.Temperature Field Measurement and Analysis of High⁃pressure Critical Flow Venturi Nozzle Gas Flow Standard Device[J]. Acta Metrologica Sinica202546(3):368-375.
[3]
刘彦军, 张毅治, 张永胜, 等. 燃油流量计在非稳态下计量性能研究[J]. 计测技术201939(1): 38-41.
[4]
吴建平, 徐科军, 许伟, 等. 基于电压电流比的瞬态电磁流量测量方法[J]. 计量学报201940(5): 829-834
WU J P XU K J XU W, et al. Transient Electromagnetic Flow Measurement Method Based on Voltage-to-current Ratio and its Realization[J]. Acta Metrologica Sinica201940(5):829-834.
[5]
MAMORU O HISASHI U KAICHIRO M, et al. CHF in oscillatory flow boiling channels[J]. Chemical Engineering Research and Design200179(4): 389-401.
[6]
赵大卫, 苏光辉, 张友佳, 等. 流量正弦脉动下圆管内单相换热的实验研究[J]. 核动力工程201031(S1): 69-72.
ZHAO D W SU G HE ZHANG Y J, et al. Experimental Study on Single-Phase Heat Transfer in Circular Tube under Oscillatory Flow Conditions[J]. Nuclear Power Engineering201031(S1): 69-72.
[7]
张永胜, 张毅治, 王鹏. 孔板与文丘里管脉动流量测量的误差比较[J]. 计量技术2020(2):63-65.
[8]
ISO/TR3313 Measurement of fluid flow in closed conduits-Guidelines on the effects of flow pulsations on flow measurement instruments [S]. 2018.
[9]
冯俊学. 用于动态流量测量的压差孔板式流量计研究[D].秦皇岛:燕山大学, 2013.
[10]
张永胜, 张毅治, 王鹏. 取压口位置对文丘里管脉动流量测量的影响[J]. 计测技术201939(3): 69 -72.
ZHANG Y S ZHANG Y Z WANG P. Influence of Pressure Tapping Position on the Measurement of Pulsating Flow Through Venturi Tube[J]. Metrology&Measurement Technology201939(3): 69 -72.
[11]
苏彦勋, 杨有涛. 流量检测技术[M]. 北京: 中国质检出版社, 2012.
[12]
王池, 王自和. 流量测量技术全书[M]. 北京: 化学工业出版社, 2012.
[13]
用安装在圆形截面管道中差压装置测量满管流体流量—第4部分:文丘里管:GB/T 2624.4 [S]. 2006.
[14]
ISO 5167-1 Measurement of fluid flow by means of pressure differential devices inserted in circular cross-section conduits running full—Part 1: General principles and requirements [S]. 2022.
[15]
朱云. 脉动流对差压式流量计测量误差影响的研究[J]. 仪器仪表学报200627(8):894-897.
ZHU Y. Study on the measurement error of differential pressure flowmeter in pulsating flow[J]. Chinese Journal of Scientific Instrument200627(8):894-897.
[16]
宋文伟,纪纲. 流动脉动对流量计测量仪表影响的几个实例[J]. 石油化工自动化20123(5):75-79.
SONG W W JI G. The Examples as the Effect of Flow Pulsation on Flow Measurement Instrumentation[J]. Automation In Petro-Chemical Industry20123(5):75-79.
[17]
MAJID N KAMRAN S. Acritical review on advanced velocity measurement techniques in pulsating flows[J]. Measurement Science and Technology201021(4): 1-19.
[18]
GOLTSMAN A SAUSHIN I MIKHEEV N, et al. Generation of sinusoidal pulsating flows in the channels of experimental setups[J]. Flow Measur Instrum201966: 60–66.
[19]
何军山, 赵玲. 差压流量计测量脉动流综述[C]. 中国核学会核能动力学会反应堆热工流体专业委员会. 第十届全国反应堆热工流体力学会议. 北京, 2007:226-230.
[20]
陈珙, 张宏建, 李海青. 引压管路对动态差压检测影响的分析[J]. 计量学报199920(4): 286-291.
CHEN G ZHANG H J LI H Q. The Effect of Pressure Ducts on Dynamic Characteristics of Pressure Drop Measuring System[J]. Acta Metrologica Sinica199920(4): 286-291.
PDF(3718 KB)

Accesses

Citation

Detail

Sections
Recommended

/