Periodicity and Symmetry in Quantization Error of Phase Angle of Sinusoidal Waveform Generated by Digital-to-Analog Converter |
LU Zu-liang |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract A sinusoidal waveform generated by digital-to-analog converter is actually a fundamental of staircase waveform. The quantization error of phase angle of this kind of the sinusoidal waveform is defined. Its intrinsic characteristics such as periodicity and symmetry are described and proved. The relative zero-point distribution is derived. The application of these characteristics in building of phase angle with higher accuracy is discussed. This zero-point distribution is quantization too with a width of 0.5H (H=2π/N) or 0.25H for odd numbers of N. In order to overcome the quantization error of phase angle, it is advised to the phase angler become an integer multiple, p, of this width, where parameters of p and N can be adjusted to such that an arbitrarily phase angle difference demand can be realized in principle within a specified uncertainty.
|
Received: 11 March 2021
Published: 15 July 2021
|
|
|
|
|
[1]Turgel R S, Oldham N M. High-precision audio-frequency phase calibration standard [J]. IEEE Trans Instrum Meas, 1978, 27: 460-464.
[2]Hess D T, Clarke K K. Phase measurement, traceability, and verification theory and practice [J]. IEEE Trans Instrum Meas, 1990, 39 (1): 52-55.
[3]Kawagoe J, Kawasaki T. A New Precision Digital Phase Meter and Its Simple Calibration Method [J]. IEEE Trans Instrum Meas, 2010, 59 (2): 396-403.
[4]Oldham N M, Turgel R S. A Power Factor Standard Using Digital Waveform Generator [J]. IEEE Trans Power Apparatus and Systems, 1981, PAS-100 (11): 4435-4438.
[5]Kinard J R, Harris L A. Wattmeter Calibration at Zero Power Factor Using Digitally Generated Sinewaves [J]. IEEE Trans Instrum Meas, 1976, 25 (4): 547-549.
[6]陆祖良, 王磊, 李敏, 等, 基于非整周期采样的工频谐波精密测量研究[J]. 计量学报, 2007, 28(4A): 5-9.
Lu Z L, Wang L, Li M, et al. Precision Measurement of Harmonics at Industrial Frequency Based on the Non-integer-period Sampling Approach [J]. Acta Metrologica Sinica, 2007, 28(4A): 5-9.
[7]Zhang J T, Pan X L. AC Power Standard at Frequencies Up to 100 kHz [C]//CPEM, 2018.
[8]贾正森, 王磊, 徐熙彤, 等, 基于约瑟夫森量子电压的交流功率测量系统及方法研究[J]. 计量学报, 2020, 41(4): 469-474.
Jia Z S, Wang L, Xu X T, et al. Research on AC Power Measurement System and Method Based on Josephson Quantum Voltage [J]. Acta Metrologica Sinica, 2020, 41(4): 469-474.
[9]Budovsky I. Measurement of Phase Angle Errors of Precision Current Shunts in the Frequency Range from 40 Hz to 200 kHz [J]. IEEE Trans Instrum Meas, 2007, 56 (2): 284-288.
[10]Pan X L, Zhang J T, Ma X F, et al. A coaxial time constant standard for the determination of phase angle errors of current shunts [J]. IEEE Trans Instrum Meas, 2013, 62 (1): 199-204.
[11]杨雁, 黄璐, 王伟, 等. NIM新一代二端对电容电桥装置[J]. 计量学报, 2020, 41(3): 284-289.
Yang Y, Huang L, Wang W, et al. The New Two Terminal Pair Capacitance Bridge at NIM [J]. Acta Metrologica Sinica, 2020, 41(3): 284-289.
[12]谷静, 杨雁, 陆青, 等, 基于数字比例技术的高精度交流电桥研究[J]. 仪器仪表学报, 2020, 41(7): 29-37.
Gu J, Yang Y, Lu Q, et al. Research on high precision AC bridge based on digital ratio technique [J]. Chinese Journal of Scientific Instrument, 2020, 41(7): 29-37.
[13]Wang Y C, Schlamminger S, Waltrip B, et al. Evaluations of a Sampling Impedance Bridge [C]//CPEM, 2020.
[14]Lee J, Schurr J, Nissila J, et al. Programmable Josephson arrays for impedance measurements [J]. IEEE Trans Instrum Meas, 2011 60(7): 2596–26011.
[15]Overney F, Flowers-Jacobs N E, Jeanneret B, et al. Josephson-based full digital bridge for high-accuracy impedance comparisons [J]. Metrologia, 2016, 53(4): 1045.
[16]Zhang Z H, Wang D G, Hu Z, et al. A precise measurement of QHR in NIM [J]. IEEE Trans on Instrum & Meas, 1991, 40: 889-892.
[17]陆祖良, 黄璐, 杨雁. 数字模拟转换器阶梯波量子误差补偿 [J]. 计量学报, 2012, 33 (3): 249-254.
Lu Z L, Huang L, Yang Y. Compensation of Quantization Error in Staircase Waveform Generated by DAC [J]. Acta Metrologica Sinica, 2012, 33(3): 249-254.
[18]陆祖良, 杨雁, 黄璐, 等. 阶梯波性质的进一步探讨——阶梯波研究之一[J]. 计量学报, 2018, 39 (6): 759-767.
Lu Z L, Yang Y, Huang L, et al. Further Discussion on Characteristics of Staircase Waveform[J]. Acta Metrologica Sinica, 2018, 39 (6): 759-767.
[19]陆祖良, 杨雁, 黄璐, 等. 正弦波电压差分测量及多周期策略——阶梯波研究之三[J]. 计量学报, 2019, 40 (2): 319-328.
Lu Z L, Yang Y, Huang L, et al. Differential Measurement of Sine Wave Voltage and Multi-period Strategy[J]. Acta Metrologica Sinica, 2019, 40 (2): 319-328.
[20]陆祖良, 杨雁, 黄璐, 等. 基于阶梯波的周期非正弦电压精密测量——阶梯波研究之四[J]. 计量学报, 2019, 40 (3): 481-490.
Lu Z L, Yang Y, Huang L, et al. Precision Measurement of Non-Sinusoidal Voltage Based on Staircase Waveform[J]. Acta Metrologica Sinica, 2019, 40 (3): 481-490. |
|
|
|