Research on Lead Equivalent Measurement of Narrow X-ray Beam Shielding Material |
HAN Lu1,2,ZHAO Rui2,WU Jin-jie2,LU Chun-hai1,WEN Yu-qin1,2,YAN Yong-qiang2,3 |
1. Chengdu University of Technology, Chengdu, Sichuan 610059, China
2. National Institute of Metrology, Beijing 100029, China
3. China Jiliang University, Hangzhou, Zhejiang 310018, China |
|
|
Abstract The shielding performance of radiation protection materials is usually expressed in lead equivalent. In order to measure the shielding ability of different protective materials accurately, according to the IEC 61331 standard, the X-ray radiation quality for lead equivalent measurement has been researched and established. The radiation quality spectrum was simulated by using EGSnrc software to study the influence of different thicknesses of lead on the radiation quality spectrum. The narrow beam condition measurement method of X-ray lead equivalent has been discussed, and the lead equivalent of 16 different protective materials under narrow beam condition was tested. The results showed that the established radiation quality meets the standard requirements. Under the condition of 120kV tube voltage, the lead equivalent measurement result of the protective material is equal to or greater than the nominal lead equivalent value, indicating that the shielding effect of the tested material meets the requirements.
|
Received: 29 October 2020
Published: 23 March 2021
|
|
|
|
|
[1]Yaffe M J, Mawdsley G E, Lilley M, et al. Composite Materials for X-ray Protection [J]. Health Physics, 1991, 60(5): 661-664.
[2]Adliene D, Gilys L, Griskonis E. Development and characterization of new tungsten and tantalum containing composites for radiation shielding in medicine [J]. Elsevier BV, 2020, 467: 21-26.
[3]Jones A K, Wagner L K. On the (f)utility of measuring the lead equivalence of protective garments [J]. Medical Physics, 2013, 40(6): 063902.
[4]Hubbert T E, Vucich J J, Armstrong M R. Lightweight aprons for protection against scattered radiation during fluoroscopy [J]. AJR Am J Roentgenol, 1993, 161(5): 1079-1081.
[5]Christodoulou E G, Goodsitt M M, Larson S C, et al. Evaluation of the transmitted exposure through lead equivalent aprons used in a radiology department, including the contribution from backscatter [J]. Medical Physics, 2003, 30(6): 1032-1038.
[6]Schlattl H, Zankl M, Eder H, et al. Shielding properties of lead-free protective clothing and their impact on radiation doses [J]. Medical Physics, 2007, 34(11): 4269-4280.
[7]Li Q X, Wei Q L, Zheng W J, et al. Enhanced Radiation Shielding with Conformal Light-Weight Nanoparticle-Polymer Composite [J].
ACS Appl Mater Interfaces, 2018,10(41):35510-35515.
[8]赵士庵, 欧向明. 射线屏蔽材料在防护中的铅当量[J]. 中华放射医学与防护杂志, 1996, (03): 211.
Zhao S A, Ou X M. Lead equivalent of radiation shielding materials in protection[J]. Chinese Journal of Radiological Medicine and Protection, 1996, (03): 211.
[9]IEC 61331-1:2014Protective Devices Against Diagnostic Medical X-Radiation-Part1:Determination of Attenuation Properties of Materials
[S].
[10]葛双, 吴金杰, 邱小平, 等. 工业钼靶X射线光机能谱的蒙特卡罗模拟 [J]. 计量学报, 2015, 36(6): 647-651.
Ge S, Wu J J, Qiu X P, et al. Monte Carlo simulation of X-ray spectroscopy of industrial molybdenum target [J]. Acta Metrologica Sinica, 2015, 36(6): 647-651.
[11]吕雅竹, 郭彬, 李德红, 等. 诊断X射线能谱影响因素的模拟与分析 [J]. 计量学报, 2020, 41(8): 1023-1026.
Lü Y Z, Guo B, Li D H, et al. Simulation and analysis of influencing factors of diagnostic X-ray energy spectrum [J]. Acta Metrologica Sinica, 2020, 41(8): 1023-1026.
[12]Mccaffrey J P, Shen H, Downton B, et al. Radiation attenuation by lead and nonlead materials used in radiation shielding garments [J]. Medical Physics, 2007, 34(2): 530-537.
[13]杜海燕, 李凡, 吴金杰, 等. 同步辐射单能X射线空气质量衰减系数的测量 [J]. 计量学报, 2019, 40(2): 333-336.
Du H Y, Li F, Wu J J, et al. Measurement of air quality attenuation coefficient of synchrotron radiation single energy X-rays [J]. Acta Metrologica Sinica, 2019, 40(2): 333-336.
[14]Buermann L. A New Method to Measure Shielding Properties of Protective Clothing Materials [M]. Berlin Heidelberg:Springer, 2009.
[15]Schopf T, Pichler T. Radiation Protection Clothing in X-Ray Diagnostics-Influence of the Different Methods of Measurement on the Lead Equivalent and the Required Mass [J]. Fortschr Rontgenstr, 2016, 188(8): 768-775.
[16]Eder H, Schlattl H. IEC 61331-1: A new setup for testing lead free X-ray protective clothing [J]. Physica Medica, 2018, 45: 6-11. |
|
|
|