The Test Facility of the Critical Back-pressure Ratio for Small Sonic Nozzle |
ZHOU Li-yuan1,2,LI Chun-hui2,XIE Dai-liang1 |
1. College of Engineering Measurement, China Jiliang University, Hangzhou, Zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract The back pressure ratio of sonic nozzle at work must be lower than critical back-pressure ratio to realize and maintain the critical flow for the sonic nozzle. Therefore, the back pressure ratio of the sonic nozzle when the Reynolds number Red of throat is less than 2×105 should be no bigger than 0.25, or is subjected to make the actual test, which is suggested in ISO 9300. A test facility was built, which can be used to measure the critical back-pressure ratio for the sonic nozzle with throat diameter of 0.3~3mm, while the flow range of sonic nozzle can be 0.05~5m3/h. The expanded uncertainty is 0.56% (k=2).
|
Received: 11 July 2018
Published: 19 December 2019
|
|
Corresponding Authors:
Chunhui Li
E-mail: lich@nim.ac.cn
|
|
|
|
[1]蔡武昌. 微小流量仪表性能及其应用[J]. 石油化工自动化, 2009, 4(4): 1-5.
Cai W C. The Performance and Application of the Minor Flowrate Meter[J]. Automation in Petro-Chemical Industry, 2009, 4(4): 1-5.
[2]常光玲. 大气采样器计量标准的测量不确定度[J]. 化学分析计量, 2008, 17(4): 16-17.
Chang G L. Measurement Uncertainty of the Meterage Standard for the Air Sampler[J]. Chemical Analysis & Meterage, 2008, 17(4): 16-17.
[3]曹培娟, 李春辉, 崔骊水, 等. 高压pVTt法气体流量标准装置不确定度实现及验证[J]. 计量学报, 2017, 38(6): 697-701.
Cao P J, Li C H, Cui L S, et al. The Verification on the Capability of High Pressure pVTt Standard Facility[J]. Acta Metrologica Sinica, 2017, 38(6): 697-701.
[4]张翰, 李春辉, 崔骊水, 等. 高压环道气体流量标准装置[J]. 计量学报, 2017, 38(3): 324-327.
Zhang H, Li C H, Cui L S, et al. The Close Loop Standard Facility of High Pressure Gas Flow[J]. Acta Metrologica Sinica, 2017, 38(3): 324-327.
[5]崔骊水, 李鹏, 邱丽荣, 等. 微风速标准装置的建立和热线风速仪校准方法的实验研究[J]. 计量学报, 2018, 39(3): 289-293.
Cui L S, Li P, Qiu L R, et al. Experiment Investigation on Calibration of Hot Wire Anemometer Based on Low Air Speed Reference Facility[J]. Acta Metrologica Sinica, 2018, 39(3): 289-293.
[6]Nakao S I, Takamoto M. Chocking phenomena of sonic nozzles at low Reynolds numbers [J]. Flow Measurement and Instrumentation, 2000, 11(4): 285-291.
[7]Park K A, Choi Y M. The evaluation of critical pressure ratio of sonic nozzles at low Reynolds numbers[J]. Flow Measurement and Instrumentation, 2001, 12(1): 37-41.
[8]胡鹤鸣, 崔骊水, 王池. 低雷诺数音速喷嘴临界背压比的实验研究[J]. 仪器仪表学报, 2012, 33(4): 737-742.
Hu H M, Cui L S, Wang C. Experimental study on critical back-pressure ratio of sonic nozzle at low Re[J]. Chinese Journal of Scientific Instrument, 2012, 33(4): 737-742.
[9]ISO 9300, Measurement of Gas Flow by Means of Critical Flow Venturi Nozzles[S]. 2005. |
|
|
|