|
|
Research on Automatic Calibrating System for Pipette Capacity |
ZHAO Yu-xiao1,XIE Xuan-da1,SUN Bin1,ZHANG Jing-yue2,WEN Ying-jie3 |
1. College of Modern Science and Technology, China Jiliang University, Hangzhou, Zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China
3. Hangzhou Institute of Calibration and Testing for Quality and Technical Supervision, Hangzhou, Zhejiang 310019, China |
|
|
Abstract Aim at the shortcomings of manual calibration such as instability of the press, temperature deviation and data that can not been processed in batches, an automatic calibration method for pipette capacity based on simulated manual press was proposed. By measuring the pressure at the two stop points of the button of the pipette, the approximate pressing distance can be obtained; comparing and analyzing data with manual operation, the optimum press distance of simulated manual press was obtained. An automatic device for simultaneous calibration of 4 pipettes was developed. Those works were intelligently controlled including extraction, discharge, weighing, and data processing adopting manipulator and LabVIEW software, fully-automatic calibrating were realized. Research results show that comparing to manual operation, the repeatability reduces 0.0%~0.9%, and the relative error decreases by 0.1%~0.7%.
|
Received: 24 January 2018
Published: 19 April 2019
|
|
|
|
|
[1]佟林, 王金涛, 刘子勇, 等. 100μL移液器国际比对不确定度的分析与评定[J]. 计量技术, 2013, (8): 45-49.
Tong L, Wang J T, Liu Z Y, et al. Uncertainty Analysis and Evaluation for National Comparison of 100 μL Locomotive Pipette [J]. Measurement Technique, 2013, (8): 45-49.
[2]国家质量监督检验检疫总局. JJG646-2006移液器检定规程[S]. 北京:中国计量出版社, 2006.
[3]Andrew Alliance. Meet the Robot that Uses Conventional Pipettes [EB/OL]. [2017-10-18]. http://www.labmanager.com/laboratory-technology/2015/04/meet-the-robot-that-uses-conventional-pipettes#. WTOQDOz7wtQ.
[4]Andrew Alliance. Better Pipetting: Better Data, Better Science [EB/OL]. [2017-10-20]. http://www.labmanager.com/products-in-action/2015/07/better-pipetting-better-data-better-science.
[5]Curtis R H, Rundell A E. PIPETTE CALIBRATION SYSTEM: US5298978(A) [P]. 1994-03-29.
[6]Almeida N, Batista E, Filipe E. Performance studies in micropipette calibration [C]// 16th International Congress of Metrology. 2013: 06010.
[7]佟林, 王金涛, 刘子勇, 等. 100 μL微量移液器国际关键比对的数学模型研究[J]. 计量学报, 2015, 36(2): 154-157.
Tong L, Wang J T, Liu Z Y, et al. The research of mathematical models in BIPM / CIPM key comparison CCM.FF-K4.2.2011 volume comparison at 100 μL [J]. Acta Metrologica Sinica, 2015, 36(2): 154-157.
[8]费业泰. 误差理论与数据处理[M]. 7版. 北京:机械工业出版社, 2015.
[9]Lorefice S.Traceability and uncertainty a: alysis in volume measurements [J]. Measurement, 2009, 42(10): 1510-1515.
[10]国家质量监督检验检疫总局. JJF1059.1-2012 测量不确定度评定与表示[S]. 北京:中国质检出版社, 2012. |
|
|
|