|
|
Measurement of Air Kerma of Tungsten Target X-ray Tube with Molybdenum Filter |
WANG Ji1,2,LI Hong-jie1,WU Jin-jie2,REN Shi-wei1 |
1. Hebei University of Science and Technology, Shijiazhuang, Hebei 050018, China
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract In the low-energy X-ray reference radiation field of China Institute of Metrology, a tungsten target molybdenum filter X-ray radiation with a tube voltage of 23, 25, 28, 30, 35kV was established, and the additional filtration was 0.06mm Mo film. Its half-value layer measurement shows that the half-value layer is maintained at 0.346 to 0.405mm of aluminum. The effective energy is maintained at 15.4 to 16.4keV, which is significantly higher than the molybdenum target X-rays excited by the same tube voltage. The correction factor of the mammography X-ray free air ionization chamber was measured, and the measurement of the air kerma was completed.
|
|
|
|
|
|
[1]张保宁, 陈万青, 张希,等. 中国乳腺癌防控形势面临挑战[J].中华肿瘤杂志, 2016, 38(10): 798-800.
[2]陈谦谦, 薛恩生. 超声、钼靶X线联合MRI在乳腺癌术前评价中的应用[J]. 中国医学影像学杂志, 2012, 20(2): 148-152.
[3]葛双. (20~300)kV X射线参考辐射场的建立和研究[D].衡阳:南华大学,2016.
[4]吴金杰, 杨元第. 中能X射线空气比释动能的绝对测量[J].仪器仪表学报,2010,31(8):240-244.
[5]韩奎初. 测量照射量(空气比释动能)的自由空气电离室[J].核标准计量与质量, 1995, (1): 38-40.
[6]Kessler C, Burns D T, Roger P, et al. Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays[J].Metrologia, 2016 (53): 6-11.
[7]李论. X射线自由空气电离室修正因子的研究[D].成都:成都理工大学, 2016.
[8]Rogers D W O. Low energy electron transport with EGS[J].Nuclear Inst & Methods in Physics Research A, 1984, 227 (3): 535-548. |
|
|
|