|
|
Key Technology Research on Model Eye Device for Non-contact Tonometer |
ZHANG Jin-ming,ZHANG Zhong-li,WANG Can |
Shanghai Institute of Measurement and Testing Technology, Shanghai 201203, China |
|
|
Abstract To solve the value traceability of non-contact tonometer, through deeply analyzing of the Imber-Fick and Glodmann tonometer principle that involved in the intraocular pressure testing, a model eye device for non-contact tonometer is independently designed and developed, where the composite structure of the cornea and sclera, and the system correction method are adopted, the same corneal deformation and the inner pre-set pressure are controlled both during the testing and tracing, the external standard value of intraocular pressure are traced back to the quality and length, using a L-shaped torque balance. The experimental results show that the device has good repeatability and stability, where the resolution is up to 1 Pa, and pressure is continuously adjustable. The expanded uncertainty reaches at 0.08 kPa (k=2).
|
Received: 30 August 2017
Published: 12 April 2018
|
|
|
|
|
[1]Grolman B. A New Tonometer System [J]. American Journal of Optometry & Archives of American Academy of Optometry,1972,49(8):646-60.
[2]方典典.用于测量眼内压的非接触式眼压测量装置的研制[D]. 重庆:重庆大学,2012.
[3]许奕如,唐玉花,王佐祥,等.非接触式眼压计与压陷式眼压计的临床应用效果分析[J].中国实用护理杂志,2010,26(13):50-51.
[4]曾阳发,刘杏,黄晶晶,等.中央角膜厚度对Goldmann眼压计与非接触式眼压计测量值的影响[J].中国实用眼科杂志,2007,25(4):393-396.
[5]石晶明,蒋幼芹.非接触眼压计测量的评价与Goldmann眼压计的比较[J].中国实用眼科杂志,2002,20(5):370-372.
[6]吴素虹,明月,陈绮华,等.非接触式眼压计测量结果的影响因素分析与护理干预[J].中华护理杂志,2003,38(6):459-460.
[7]郭萌,陈剑,刘珏,等.LASIK术后非接触式眼压计测量值的变化[J].广东医学,2009,30(11):1642-1643.
[8]Reichert, Inc. Tonometer calibration tool: US201403 2843A1[P]. 2014-10-30.
[9]四川中测辐射科技有限公司. 非接触式眼压计校准用模拟人眼装置:CN2010205159331[P]. 2011-05-04.
[10]王莉茹,马振亚. 验光机用客观式模拟眼的研制及国际比对[J]. 计量学报,2001,22(1):23-28.
[11]徐渊,赵炜,杜红俊,等.正常视力人群眼压和角膜中央厚度的关系[J].国际眼科杂志,2007,7(4):1034-1035.
[12]张扬,赵家良,卞爱玲,等.中央角膜厚度、角膜曲率对Goldmann压平眼压计和非接触眼压计测量结果的影响[J].中华眼科杂志,2009,45(8):713-718.
[13]杨浩.人眼球物理建模及结合青光眼的生物力学分析初探[D]. 厦门:厦门大学,2006.
[14]Goldmann V H, Schmidt T H. ber Applanations Tonometrie[J]. Opthalmologica, 1957,134(4):221-242.
[15]谢杰镇.国人虚拟眼角膜建模与仿真研究[D].厦门:厦门大学,2007.
[16]张学勇.眼角膜生物力学性能非破坏性检测技术研究[D]. 合肥:合肥工业大学,2012.
[17]阎启昌,贺玉华,柏全豪,等.Goldmann压平眼压计与非接触式眼压计测量眼压的对比研究[J]. 国际眼科杂志,2006,6(3):12l-122.
[18]张彬,王大博,韩静.非接触式眼压计与Goldmann压平式眼压计测量值比较研究[J].中国实用眼科杂志,2013,31(3):270-272. |
|
|
|