Acta Metrologica Sinica  2018, Vol. 39 Issue (3): 289-293    DOI: 10.3969/j.issn.1000-1158.2018.03.01
Current Issue | Archive | Adv Search |
Experiment Investigation on Calibration of Hot Wire Anemometer Based on Low Air Speed Reference Facility
CUI Li-shui1,2,LI Peng2,QIU Li-rong1,LI Jian-shuang2
1. Beijing Institute of Technology, Beijing 100081, China 
2. National Institute of Metrology, Beijing 100029, China
Download: PDF (973 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  The low air speed reference facility is established at NIM of China, based on the device, the experiment on calibration of hot wire anemometer is investigated.The facility is described in detail and the calibration method for hot wire sensor is introduced. Based on the reciprocity principle, the sensor probe is mounted on the carriage that slides with consistent speed and the sliding stroke is measured by laser interferometer accurately during calibration process. The velocity measurement is therefore traceable to the measurement of length and time primary standards, additionally, the parameters of static equation is obtained according to theoretical module of hot wire sensor, the experimental results are consistent with theoretical equation. The experimental results and uncertainty analysis show that the expanded uncertainty of the breeze standard device is U=0.82%(k=2), and the expanded uncertainty of calibration result of hot wire anemometer is U=2.42%(k=2).
Key wordsmetrology      hot wire anemometer      micro air speed      laser interferometer      calibration      uncertainty     
Received: 18 October 2016      Published: 12 April 2018
PACS:  TB937  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
CUI Li-shui
LI Peng
QIU Li-rong
LI Jian-shuang
Cite this article:   
CUI Li-shui,LI Peng,QIU Li-rong, et al. Experiment Investigation on Calibration of Hot Wire Anemometer Based on Low Air Speed Reference Facility[J]. Acta Metrologica Sinica, 2018, 39(3): 289-293.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2018.03.01     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2018/V39/I3/289
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech