Abstract To observe the variation of matrix effect during the metrological verification of the electrolyte analyzer, the electrolyte analyzer using the calibration solution of the aqueous solution matrix, and the biochemical analyzer using the calibration solution of the human serum matrix was calibrated, and then the standard materials of aqueous solution matrixes and human serum matrixes with different concentrations was respectively determined, the significance of the verification results will be analyzed.The test results of K+, Na+, and Cl- are statistically significant (p<0.01) within certain concentration ranges after the calibration of the biochemical analyzer using human serum matrix and electrolyte analyzer using the aqueous solution matrix.Therefore, the deviation caused by the matrix effect varies with the concentration during the calibration process.
|
Received: 08 October 2016
Published: 29 December 2017
|
|
|
|
|
[1]周新, 涂植光. 临床生物化学和生物化学检验[M]. 3版. 北京:人民卫生出版社, 2006:207-208.
[2]湛玉良, 李静, 井新辉. 校准基质对测定结果的影响[J]. 陕西医学检验, 2002, 17(1):18-19.
[3]段满乐.现代临床实验室管理学[M].北京:中国科学技术出版社,2005:192-198.
[4]国家质量监督检验检疫总局.JJG 1051—2009 电解质分析仪检定规程[S].
[5]李佳乐, 武利庆, 金有训,等. 人血清白蛋白纯品含量的同位素稀释质谱方法研究[J]. 计量学报, 2016, 37(3):328-332.
[6]张玲, 盛灵慧, 王晶,等. 转基因棉花Mon15985质粒标准物质多家验证实验研究及不确定度分析[J]. 计量学报, 2013, 34(6):597-601.
[7]龙琳娟, 周月平, 徐梅. 血气分析仪、电解质分析仪与全自动生化分析仪三者所测电解质的比对分析[J]. 实验与检验医学, 2009, 27(5):501-502.
[8]吕赛平, 伍志杰, 邹学森. 基质效应的评价[J]. 现代检验医学杂志, 2008, 23(5):113-114. |
|
|
|