Technical Research on Calibration of Fully Automated ELISA Analyzers |
KANG Juan1,WU Li-qing2,YAN Zhen-yu1,WANG Jun1,SUN Jing-sheng1 |
1.Beijing Institute of Medical Device Testing & Beijing Key Lab of Medical Device Testing & Safety Evaluation, Beijing 101111, China
2. Division of Medical and Biological Measurement, National Institute of Metrology, Beijing 100029, China |
|
|
Abstract A calibration technique has been established to evaluate the testing performance of fully automated ELISA analyzers. The technique tests the accuracy and precision of adding module, the temperature of incubation module, the washing residue rate and cross contamination of washing module, and the absorbance stability, accuracy, precision and channel difference of reading module separately around the constituent modules of the system. 3 fully automated ELISA analyzers from different manufactures were chosen to validate the technique applicability. Results showed that the maximum adding bias and CV were -7.4% and 8.0% respectively, the incubation temperature bias was within ±0.5 ℃, the maximum washing residue rate was 0.8% and maximum cross contamination was 0.1%, the maximum absorbance fluctuation was 0.003 within ten minutes, the maximum absorbance bias and CV were 0.019 and 0.2% respectively for tested wavelengths and absorbance,the maximum channel difference was 0.005. The calibration technique was testified to be scientific and practical, and expected to be used to evaluate the fully automated ELISA analyzers effectively.
|
Received: 03 November 2016
Published: 27 September 2017
|
|
|
|
|
[1]陈叙. 基于免疫酶技术的全自动酶免分析系统的设计与开发 [D]. 苏州: 苏州大学, 2006.
[2]丁卫平, 柏亮.全自动加样仪与全自动酶免仪的2个关键点监控[J].中国输血杂志,2009,22(6):476-477.
[3]郑怀竞.必须做好各级血站实验室免疫学检验的室内全程质量控制[J].中国输血杂志,2002,15(5):299-300.
[4]国家质量监督检验检疫总局. JJG 861—2007 酶标分析仪检定规程 [S]. 2007.
[5]国家食品药品监督管理局. YY/T 0654—2008 全自动生化分析仪[S]. 2008.
[6]武利庆, 杨彬, 丁峰元,等. 基于吸光度测定的洗板机校准方法及其不确定度评定[J].计量技术,2012,(5):46-48.
[7]袁红, 毛旖, 黄文芳. ELISA试验加样误差对试验结果的影响分析[J].国际检验医学杂志,2009,30(9):835-837. |
|
|
|