A Miniaturized Test Device to Measure Key Resolution Parameters of an Ophthalmic Optical Coherence Tomography Equipment |
FU Xiao-yu1,HU Zhi-xiong2,GE Chun-feng1,LIU Wen-li2,HONG Bao-yu2,JIA Dong-fang1 |
|
|
Abstract Both the axial resolution and the lateral resolution were derived from a theoretical model of optical coherence tomography technology. A miniaturized test device for measuring the light source spectrum and the numerical aperture of an optical coherence tomography system is developed to verify a commercial clinical ophthalmic optical coherence tomography equipment. The acquired axial resolution is 6.9μm, while the lateral resolution has the value of 42.3μm. The preliminary results have demonstrated that the integrated testing system could be utilized as a metrological standard device for quality control of ophthalmic OCT equipment.
|
Received: 13 November 2015
Published: 27 September 2017
|
|
|
|
|
[1]Huang D, Swanson E, Lin C, et al. Optical coherence tomography[J]. Science, 1991, 254(5035): 1178-1181.
[2]孙延奎. 光学相干层析医学图像处理及其应用[J]. 光学 精密工程, 2014, 22(4): 1086-1104.
[3]王甜甜, 余晓锷. CT系统空间分辨率测量技术的改进[J]. 计量学报, 2012, 33(2): 187-192.
[4]胡志雄, 郝冰涛, 刘文丽, 等. 用于光学相干层析成像设备点扩散函数测量的模体制作与使用方法研究[J]. 光学学报, 2015, 35(4):273-279.
[5]胡志雄, 刘文丽, 洪宝玉, 等. 光学相干层析成像三维分辨率测试模拟眼[J]. 光电工程, 2014, 41(12): 28-32.
[6]Fujimoto J. Optical coherence tomography for ultrahigh resolution in vivo imaging[J]. Nature Biotechnology, 2003, 21(11): 1361-1367.
[7]唐弢, 赵晨, 陈志彦, 等. 超高分辨光学相干层析成像技术与材料检测应用[J]. 物理学报, 2015, 64(17): 174201.
[8]Boppart S A, Bouma B E, Pitris C, et al. In vivo cellular optical coherence tomography imaging[J]. Nature Medicine, 1998, 4(7): 861-865.
[9]吴彤, 丁志华. 20kHz扫频光学相干层析系统[J]. 中国激光, 2009, 36(2): 503-508.
[10]郭昕, 王向朝, 步鹏, 等. 样品散射对频域光学相干层析成像光谱形状和深度分辨率的影响[J]. 光学学报, 2014, 34(1):181-186.
[11]Ceyhun A, Pascale P, Rolland J P. Estimation of longitudinal resolution in optical coherence imaging[J]. Applied Optics, 2002, 41(25): 5256-5262. |
|
|
|